医疗卫生装备2017,Vol.38Issue(7):73-76,4.DOI:10.7687/j.issn1003-8868.2017.07.073
全数字化乳腺X线成像系统空间分辨力及平均腺体剂量检测
Test spatial resolution and average glandular dose of full field digital mammography system
摘要
Abstract
Objective To test the spatial resolution and average glandular dose (AGD) of full field digital mammography system to obtain high-resolution and contrast breast X-ray images with the lowest dose.Methods The M12 breast X-ray performance test phantom was placed on the breast support center position,then line-pair card was put on the M12 phantom and fixed groove,and exposure was carried out with auto exposure control,28 kV tube voltage and common anode filtration combination so as to detect the spatial resolution.Half-value layer (HVL) was measured by Fluke TNT12000 ray detection tools,2 mm-thickness plate and aluminum sheets with different thicknesses,and then AGD was calculated accordingly.Results The spatial resolutions were all higher than 7 lp/mm and proved to meet the quality standard,which had the vertical spatial resolution being 8 lp/mm,the maximum value of the lateral spatial resolution being 10 lp/mm and the minimum value being 8 lp/mm.AGD was lower than 2 mGy and then proved qualified which was limited within 0.60 and 0.61 mGy.Conclusion The test of spatial resolution and AGD of full digital mammography system can assess the performance and radiation dose,reflect the performance of equipment,ensure image quality and lower radiation dose.关键词
全数字化乳腺X线成像系统/空间分辨力/平均腺体剂量/乳腺X线影像/分辨力/对比度Key words
full field digital mammography system/spatial resolution/average glandular dose/breast X-ray image/resolution/contrast分类
医药卫生引用本文复制引用
梁永刚,付丽媛,倪萍,陈坚,李威,刘冰川,陈建新,陈自谦..全数字化乳腺X线成像系统空间分辨力及平均腺体剂量检测[J].医疗卫生装备,2017,38(7):73-76,4.基金项目
国家重点研发计划数字诊疗装备研发重点专项子课题(2016YFC0103103) (2016YFC0103103)
福建省科技计划项目(2016I0010) (2016I0010)
南京军区重大课题(14ZX23) (14ZX23)