| 注册
首页|期刊导航|医疗卫生装备|SYSMEX HISCL5000化学分析仪检测乙肝血清标志物性能评价

SYSMEX HISCL5000化学分析仪检测乙肝血清标志物性能评价

郑齐锶 戴蕾 陶月 宁明哲

医疗卫生装备2017,Vol.38Issue(9):90-92,104,4.
医疗卫生装备2017,Vol.38Issue(9):90-92,104,4.DOI:10.7687/j.issn1003-8868.2017.09.090

SYSMEX HISCL5000化学分析仪检测乙肝血清标志物性能评价

Clinical performance evaluation of HBV serological biomarkers using SYSMEX HISCL5000 chemiluminescence analyzer

郑齐锶 1戴蕾 1陶月 1宁明哲1

作者信息

  • 1. 210008南京,南京大学医学院附属鼓楼医院检验科
  • 折叠

摘要

Abstract

Objective To assess the detection performance of SYSMEX HISCL5000 automatic chemiluminescence immunoassay analyzer for serum hepatitis B virus (HBV) biomarkers.Methods Serological HBV biomarkers,including HBsAg,HBsAb,HBeAg,HbeAb and HbcAb,were measured by HISCLS000 analyzer.Further,a panel of parameters were analyzed,including precision,cross contamination rate,linear range,concordance rate,biological reference interval and limit of detection.According to the guideline from Clinical & Laboratory Standards Institues (CLSI) EP system,the potential clinical application of using HISCL5000 analyzer to measure serum HBV biomarkers were evaluated.Results A panel of five serum HBV biomarkers was measure by using HISCLS000 analyzer.The coefficient of variation (CV) value of the within-run imprecision was from 0.60% to 4.17%,and CV value of the between-run imprecision was from 0.04% to 5.35%.Linear verification showed that r2 was between 0.980 5 and 0.998 7,and a was between 0.970 9 and 1.022 6.The ratio of cross contamination was 0.00%.The coincident rate of HISCL5000 analyzer with other methods was between 96.00% and 100.00%.Biological reference interval and limit of detectionderived from this analyzer were also proved qualified.Conclusion HISCL5000 analyzer can be used clinically to detect HBV biomarkers.

关键词

SYSMEX HISCL5000化学分析仪/乙肝两对半/血清标志物/乙肝/性能评价

Key words

SYSMEX HISCLS000 analyzer/second liver two half-and-half/serum marker/HBV/performance evaluation

分类

医药卫生

引用本文复制引用

郑齐锶,戴蕾,陶月,宁明哲..SYSMEX HISCL5000化学分析仪检测乙肝血清标志物性能评价[J].医疗卫生装备,2017,38(9):90-92,104,4.

医疗卫生装备

OACSTPCD

1003-8868

访问量0
|
下载量0
段落导航相关论文