深圳大学学报(理工版)2017,Vol.34Issue(5):532-536,5.DOI:10.3724/SP.J.1249.2017.05532
分压电路对X射线条纹相机成像稳定性的影响
Influence of voltage divider circuit on imaging stability of X-ray streak camera
摘要
Abstract
In the static test of the X-ray streak camera,it is shown that the stability of the voltage divider circuit significantly affects the imaging performance of the streak camera.In order to improve the imaging stability,we analyze the voltage divider circuit.We find that the output instability of the voltage divider circuit is caused by the current thermal effect.The voltage circuit with smaller current is adopted to reduce the current thermal effect.We test and compare the voltages,current outputs,the static imaging performances of the streak cameras and the system modulation transfer functions for the two voltage dividers.Both theoretical analysis and experimental results show that after the camera has been powered with the smaller current divider circuit for 20 min,the operating current,the system temperature and electron lens focal length change by 1.4%,7.5% and 0.14%,respectively.And the static image quality and modulation transfer function are improved.关键词
光电子学与激光技术/惯性约束聚变/核仪器/X射线条纹相机/条纹变像管/分压电路/调制传递函数Key words
optoelectronics and laser technology/inertial confinement fusion/nuclear instrumentation/X-ray streak camera/streak tube/voltage divider circuit/modulation transfer function分类
能源科技引用本文复制引用
顾礼,李翔,宗方轲,周军兰,杨勤劳,郭宝平..分压电路对X射线条纹相机成像稳定性的影响[J].深圳大学学报(理工版),2017,34(5):532-536,5.基金项目
国家重大科学仪器设备开发专项资助项目(2014YQ230659) (2014YQ230659)
国家自然科学基金资助项目(11705119) (11705119)
广东省自然科学基金项目(2017A030310142) (2017A030310142)
深圳市科技计划资助项目(JCYJ20170302152748002)National Key Scientific Instrument and Equipment Development Projects (2014YQ230659) (JCYJ20170302152748002)
National Natural Science Foundation of China (11705119) (11705119)
Natural Science Foundation of Guangdong Province (2017A030310142) (2017A030310142)
Shenzhen Science and Technology Project (JCYJ20170302152748002) (JCYJ20170302152748002)