| 注册
首页|期刊导航|中国医学装备|酶标分析仪检定结果的不确定度分析评定

酶标分析仪检定结果的不确定度分析评定

崔骊 李向东

中国医学装备2017,Vol.14Issue(11):42-44,3.
中国医学装备2017,Vol.14Issue(11):42-44,3.DOI:10.3969/J.ISSN.1672-8270.2017.11.012

酶标分析仪检定结果的不确定度分析评定

A evaluation for uncertainty analysis of the verification result of ELISA analyzer(microplate reader)

崔骊 1李向东1

作者信息

  • 1. 西京医院医学工程科 陕西 西安 710032
  • 折叠

摘要

Abstract

Objective: To analyze various factors that affect the verification result in the verification by using ELISA analyzer and assess the accuracy of verification results of ELISA analyzer. Methods: Through the comparison of verification results and the calculation for the repeatability of measurement standard, the accuracy of standard device was analyzed and verified. Results: Through analyzed the effect of repeatability of operation and standard device, the factors that affected verification results were eliminated. And the accuracy of verification results were guaranteed. Conclusion: In the process of verification for ELISA analyzer, operators should pay attention to the setting of the verification instrument. And the operation regulation can ensure the accuracy and reliability of the measured value of spectrophotometer. Besides, operators also should pay attention to the analysis of verification results and confirm whether the repeatability of the verification results can meet the requirement of practice.

关键词

酶标分析仪/不确定度/分光光度计/质量控制/评定

Key words

ELISA analyzer/Uncertainty/Spectrophotometer/Quality control/Evaluation

分类

医药卫生

引用本文复制引用

崔骊,李向东..酶标分析仪检定结果的不确定度分析评定[J].中国医学装备,2017,14(11):42-44,3.

基金项目

陕西省社会发展科技攻关项目(2016SF-062)"器材设备管理信息系统的改进研究" (2016SF-062)

中国医学装备

OACSTPCD

1672-8270

访问量4
|
下载量0
段落导航相关论文