| 注册
首页|期刊导航|计量学报|400~1700nm绝对光谱响应度校准和量值传递

400~1700nm绝对光谱响应度校准和量值传递

史学舜 徐楠 刘长明 甘海勇 李健军 林延东 陈坤峰

计量学报2017,Vol.38Issue(6):681-685,5.
计量学报2017,Vol.38Issue(6):681-685,5.DOI:10.3969/j.issn.1000-1158.2017.06.05

400~1700nm绝对光谱响应度校准和量值传递

Calibration and Value Transmission of Absolute Spectral Responsivity between 400 nm and 1700 nm

史学舜 1徐楠 2刘长明 1甘海勇 2李健军 3林延东 2陈坤峰1

作者信息

  • 1. 中国电子科技集团公司第四十一研究所,山东青岛266555
  • 2. 中国计量科学研究院,北京100029
  • 3. 中国科学院通用光学定标与表征技术重点实验室,安徽合肥230031
  • 折叠

摘要

Abstract

The principle and methods of optical radiation measurement was illustrated.The absolute spectral responsivity of Si trap detectors and InGaAs trap detectors were calibrated with the absolute cryogenic radiometer at eleven wavelengths between 400 nm and 1700 nm,which are 476.1,488,514.7,521,568,632.8,647.1,785,852,980,1064 and 1550 nm,respectively,with measurement uncertainties better than 0.05%.The absolute spectral responsivity curves of Si trap detectors were calculated by quantum efficiency model.The absolute responsivity of three wavelengths was applied to get absolute spectral responsivity from the relative spectral responsivity of InGaAs detector.Results show that the two types of trap detectors can be used to calibrate and transfer optical radiation value in the visible and near infrared spectral range.

关键词

计量学/绝对光谱响应度/绝对低温辐射计/陷阱探测器/不确定度

Key words

metrology/absolute spectral responsivity/absolute cryogenic radiometer/trap detector/uncertainty

分类

通用工业技术

引用本文复制引用

史学舜,徐楠,刘长明,甘海勇,李健军,林延东,陈坤峰..400~1700nm绝对光谱响应度校准和量值传递[J].计量学报,2017,38(6):681-685,5.

基金项目

国家“863”计划(2015AA123702) (2015AA123702)

国家技术基础项目(J312013A001) (J312013A001)

计量学报

OA北大核心CSCDCSTPCD

1000-1158

访问量0
|
下载量0
段落导航相关论文