原子能科学技术2017,Vol.51Issue(12):2326-2329,4.DOI:10.7538/yzk.2017.51.12.2326
中国先进研究堆实验室HPGe探测器
无源效率刻度方法研究
Research on Detection Efficiency of HPGe Detector without Standard Source Calibration in CARR Laboratory
张念鹏 1丁丽1
作者信息
- 1. 中国原子能科学研究院反应堆工程技术研究部,北京 102413
- 折叠
摘要
Abstract
Considering the real fact of CARR Laboratory 's HPGe detector ,the detec-tion efficiency without standard source calibration was calculated .The detection effi-ciency calculated by Monte Carlo method was compared with that by numerical integra-tion method in terms of energy and volume .After measuring the activated piece (point source) and water sample (volume source) ,the experimental results agree well with calculated results ,validating the feasibility and accuracy of calibration method without standard source .Compared with standard source calibration experiment ,this method breaks through many limitations ,reduces the work of preparing standard source ,saves lots of time ,and also greatly broadens the application scope of HPGe detector .关键词
HPGe探测器/探测效率/蒙特卡罗方法/数值积分方法/标准源验证Key words
HPGe detector/detection efficiency/Monte Carlo method/numerical inte-gration method/standard source verification分类
能源科技引用本文复制引用