四川环境2017,Vol.36Issue(6):91-95,5.
利用天然本底辐射场对固定式环境辐射监测仪进行现场校准方法的研究
Research on the On-site Calibration Method of Fixed Environmental Radiation Monitor Using Natural Background Radiation Field
王亮 1于兵 2但玉娟 2李燕1
作者信息
- 1. 四川省辐射环境管理监测中心站,成都611139
- 2. 中国测试技术研究院,成都610021
- 折叠
摘要
Abstract
In this study,we have carried out on-site calibration experiment of stationary environmental radiation monitoring instrument in natural radiation field.The results show that using on-site calibration experiment in natural radiation field can solve the problem of the calibration of stationary environmental radiation monitoring instrument In natural field,using integral method can obtained stable data and stable relative response,and has good correlation.By changing the position of exchange transfer instrument and monitoring instrument,can accurately confirm the relative response.The average of relative response is 0.996 and the standard deviation is 0.12 %.关键词
天然本底/现场/校准。Key words
Natural background/scene/calibration分类
资源环境引用本文复制引用
王亮,于兵,但玉娟,李燕..利用天然本底辐射场对固定式环境辐射监测仪进行现场校准方法的研究[J].四川环境,2017,36(6):91-95,5.