碳纤维固定底板对放射治疗射野剂量分布影响的初步研究

柏正璐 时飞跃 秦伟 蒋红兵

中国医学装备2018,Vol.15Issue(1):22-25,4.
中国医学装备2018,Vol.15Issue(1):22-25,4.DOI:10.3969/J.ISSN.1672-8270.2018.01.007

碳纤维固定底板对放射治疗射野剂量分布影响的初步研究

The primary research for the influence of immobilization baseplate of carbon fiber on dose distribution of radiation field of radiotherapy

柏正璐 1时飞跃 2秦伟 2蒋红兵1

作者信息

  • 1. 南京医科大学附属南京医院(南京市第一医院)医疗设备处 江苏 南京 210006
  • 2. 南京医科大学附属南京医院(南京市第一医院)肿瘤放疗中心 江苏 南京 210006
  • 折叠

摘要

Abstract

Objective:To analyze the attenuation of immobilization baseplate of carbon fiber for ray and the change of dose verification results for enhanced radiation field of MatriXX. And to primarily research the influence of this immobilization baseplate for the dose distribution of radiation field of radiotherapy.Methods: 20 patients with cervical cancer were enrolled in the research. The angle of gantry was 180°in the enhanced plan of immobilization field, and the corresponding verification plans were designed. When the verification plan was implemented on the Varian Clinac iX of accelerator, the two-dimensional ionization chamber array (MatriXX) was used to measure dose distributions of the plan. The immobilization baseplate of carbon fiber (ORFIT AIO) was used in the research. The situations of no baseplate (A group), one baseplate (B group), two baseplates (C group)and three baseplates (D group)between the handpiece of accelerator and Matri XX were researched, respectively. The data of the 4 groups were obtained by using software (OminiPro I'mRT), and the largest dose point and passing rate of Gamma in plane of radiation field were counted and analyzed.Results: Compared with A group, the average attenuation factors for ray on the largest dose point of B, C and D group in the plane of radiation field were (1.73±0.11)%, (3.44±0.36)%, and (5.48±0.44)%, respectively. The passing rates of Gamma in the 4 groups were (94.87±2.31)%, (99.18±0.61)%, (99.79±0.27)% and (96.24±1.99)%, respectively. And the differences of passing rate between each group and A group were statistical significance (t=9.788,t=9.261,t=2.256,P<0.05).Conclusion:The immobilization baseplate of carbon fiber has influence for the attenuation of high-energy X-rays and it can affect the dose distribution of radiation field which has passed baseplate in some degrees. And when physicists calculates the dose, they should consider the effect of immobilization baseplate of carbon fiber for dose distribution of radiation field.

关键词

碳纤维/固定底板/剂量衰减/γ射线通过率/MatriXX/调强/医用直线加速器

Key words

Carbon fiber/Immobilization baseplate/Dose attenuation/Gamma passing rate/MatriXX/Intensity modulated/Medical linear accelerator

分类

医药卫生

引用本文复制引用

柏正璐,时飞跃,秦伟,蒋红兵..碳纤维固定底板对放射治疗射野剂量分布影响的初步研究[J].中国医学装备,2018,15(1):22-25,4.

基金项目

南京市医学科技发展资金资助(QRX11033)"南京市卫生青年人才培养工程" (QRX11033)

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文