医疗卫生装备2018,Vol.39Issue(1):16-18,36,4.DOI:10.7687/j.issn1003-8868.2018.01.016
基于回归分析的尿流率计校准装置排水量模型研究
Research on displacement model of urine flow rate meter calibration device based on regression analysis
摘要
Abstract
Objective To determine the relationship between the theory and actual displacement of the urine flow rate meter calibration device gear pump, and to construct a model between actual and theory displacement, in order to obtain actual displacement via theory displacement. Methods The weight of the standard water flow in each flow rate was measured by B2000S digital scales,and the scatter plot between the theory displacement and actual displacement was drawn to find the best fitted curve;then the MATLAB Curve Fitting Tool was used to obtain the best quasi relational equation. Results The error between the fitted displacement and the actual displacement of the gear bump was 1.494% in maximum and-0.010% in minimum, far less than 5% of the requirements for urinary flow meter measurement. Conclusion Based on regression analysis, the relationship between the theoretical displacement and actual displacement of the flow rate meter calibration device is established,the fitting error is small,and the method is simple and easy to achieve.关键词
尿流率计校准装置/排水量/回归分析/MATLAB/拟合曲线Key words
urine flow rate meter calibration device/displacement/regression analysis/MATLAB/fitting curve分类
医药卫生引用本文复制引用
姬军,张金萍,张延,王莉新,高佳硕,肖宏..基于回归分析的尿流率计校准装置排水量模型研究[J].医疗卫生装备,2018,39(1):16-18,36,4.基金项目
军事医学技术基础科研项目(军装计2016[599]) (军装计2016[599])
军事医学计量专项(2012-JL1-016) (2012-JL1-016)