| 注册
首页|期刊导航|中国医学装备|基于Proteus的PX2001型野战洗手器的功能改进与仿真

基于Proteus的PX2001型野战洗手器的功能改进与仿真

张伟 李效银 尹晓峰 张谷敏 王静雅

中国医学装备2018,Vol.15Issue(4):31-33,3.
中国医学装备2018,Vol.15Issue(4):31-33,3.DOI:10.3969/J.ISSN.1672-8270.2018.04.008

基于Proteus的PX2001型野战洗手器的功能改进与仿真

The improved PX2001 type hand-washing device of field based on Proteus and its circuit simulation

张伟 1李效银 2尹晓峰 1张谷敏 1王静雅1

作者信息

  • 1. 济南军区联勤部药品仪器检验所 山东 济南 250022
  • 2. 解放军第89医院放射科 山东 潍坊 261021
  • 折叠

摘要

Abstract

Objective: To improve and upgrade the PX2001 type hand-washing device of field, and ensure its safety and reliability in works. Methods: Using HMS87C1202AP single-chip microcomputer to control the hand-washing device. And monitoring device of liquid level and acoustooptic alarm circuit were added to real-time device so as to monitor the liquid level and raise the alarm when liquid level was too low. Results: The circuit can accurately monitor the liquid level. And the circuit can be cut off automatically when the liquid level was too low, and it can forcibly protect pump device. Conclusion: The improved hand-washing device has the following advantages, included of low cost, easy-operating, safety and reliability, high stability etc., which contribute to raise the utilization rate and play a more effectively role for decontamination in both wartime and peacetime.

关键词

洗手器/HMS87C1202AP/Proteus/改进/电路仿真

Key words

Hand-washing device/HMS87c1202AP/Proteus/Technical improvement/Circuit simulation

分类

医药卫生

引用本文复制引用

张伟,李效银,尹晓峰,张谷敏,王静雅..基于Proteus的PX2001型野战洗手器的功能改进与仿真[J].中国医学装备,2018,15(4):31-33,3.