| 注册
首页|期刊导航|中国医疗设备|医科达SYNERGY直线加速器PRF EN CHK连锁维修案例

医科达SYNERGY直线加速器PRF EN CHK连锁维修案例

孙波 丛宁宁

中国医疗设备2018,Vol.33Issue(4):101-103,3.
中国医疗设备2018,Vol.33Issue(4):101-103,3.DOI:10.3969/j.issn.1674-1633.2018.04.026

医科达SYNERGY直线加速器PRF EN CHK连锁维修案例

Troubleshooting of Elekta Synergy PRF EN CHK Interlock

孙波 1丛宁宁1

作者信息

  • 1. 烟台毓璜顶医院 医学工程处,山东 烟台 264000
  • 折叠

摘要

Abstract

During the operation of elekta synergy linear accelerator, programmable pulse generator (PPG) PCB will transmit pulse repetition frequency (PRF) ENABLE signal, pass through three control areas, multileaf collimator and linac control system, then go back to PPG PCB. If everything works well, PPG PCB will provide thyristor trigger pulse to high tension area, then pulse forming network will charge and discharge normally, the magnetron will generate microwave to accelerating tube, and the linear accelerator will works properly. Otherwise, if any part works in wrong conditions, the linear accelerator will not delivery beam. During the clinical use of elekta synergy linear accelerator in our hospital, the PRF ENABLE interlock occurred several times. This article summarized the PRF EN CHK chain of elekta Synergy linear accelerator in the process of operation in our hospital, analyzed in detail the happened or possibile failures, focued on the process of troubleshooting, which might provide the reference for the related personnel.

关键词

直线加速器/脉冲重复频率/高压连锁电路/连锁排查/医疗设备维修

Key words

linear accelerator/pulse repetition frequency/high tension interlock/troubleshooting/medical equipment maintenance

分类

能源科技

引用本文复制引用

孙波,丛宁宁..医科达SYNERGY直线加速器PRF EN CHK连锁维修案例[J].中国医疗设备,2018,33(4):101-103,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文