中国医疗设备2018,Vol.33Issue(2):90-93,4.DOI:10.3969/j.issn.1674-1633.2018.02.023
不同输液管路的适用性分析
Applicability Analysis of Different Infusion Pipe
凌文嘉 1陈曼珊 1邓映雪2
作者信息
- 1. 广东省人民医院(广东省医学科学院) 机电设备维修科,广东 广州 510080
- 2. 广州市中医医院 设备科,广东 广州 510130
- 折叠
摘要
Abstract
Objective To evaluate the applicability of a variety of brands of transfusion pipe in the same model. Methods The flow rate and the blocking alarm was firstly tested via quality control of the infusion pump, then the data statistic analysis was carried out. Results The results indicated that there were some differences in the testing of different infusion pipelines. In addition, the flow velocity error was different at different velocity. Moreover, the pressure threshold and response time had exceeded the error range after long use. Conclusion To ensure the quality of clinical treatment effect, the clinical quality control personnel should target the different infusion pump brand in this hospital to choose the suitable infusion pipeline, and instruct the clinical staff to operate correctly.关键词
输液泵/质量控制/相对误差/输液管路Key words
infusion pump/quality control/the relative error/infusion pipe分类
医药卫生引用本文复制引用
凌文嘉,陈曼珊,邓映雪..不同输液管路的适用性分析[J].中国医疗设备,2018,33(2):90-93,4.