| 注册
首页|期刊导航|中国医疗设备|不同输液管路的适用性分析

不同输液管路的适用性分析

凌文嘉 陈曼珊 邓映雪

中国医疗设备2018,Vol.33Issue(2):90-93,4.
中国医疗设备2018,Vol.33Issue(2):90-93,4.DOI:10.3969/j.issn.1674-1633.2018.02.023

不同输液管路的适用性分析

Applicability Analysis of Different Infusion Pipe

凌文嘉 1陈曼珊 1邓映雪2

作者信息

  • 1. 广东省人民医院(广东省医学科学院) 机电设备维修科,广东 广州 510080
  • 2. 广州市中医医院 设备科,广东 广州 510130
  • 折叠

摘要

Abstract

Objective To evaluate the applicability of a variety of brands of transfusion pipe in the same model. Methods The flow rate and the blocking alarm was firstly tested via quality control of the infusion pump, then the data statistic analysis was carried out. Results The results indicated that there were some differences in the testing of different infusion pipelines. In addition, the flow velocity error was different at different velocity. Moreover, the pressure threshold and response time had exceeded the error range after long use. Conclusion To ensure the quality of clinical treatment effect, the clinical quality control personnel should target the different infusion pump brand in this hospital to choose the suitable infusion pipeline, and instruct the clinical staff to operate correctly.

关键词

输液泵/质量控制/相对误差/输液管路

Key words

infusion pump/quality control/the relative error/infusion pipe

分类

医药卫生

引用本文复制引用

凌文嘉,陈曼珊,邓映雪..不同输液管路的适用性分析[J].中国医疗设备,2018,33(2):90-93,4.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|