首页|期刊导航|中国医学装备|NeuViz 128型CT颈部增强图像质量与辐射剂量的比较评估

NeuViz 128型CT颈部增强图像质量与辐射剂量的比较评估

曹锋 王传彬 董江宁 高飞 李乃玉

中国医学装备2018,Vol.15Issue(5):10-13,4.
中国医学装备2018,Vol.15Issue(5):10-13,4.DOI:10.3969/J.ISSN.1672-8270.2018.05.004

NeuViz 128型CT颈部增强图像质量与辐射剂量的比较评估

Comparative study of image quality and radiation dose of NeuViz 128 CT CT on neck enhanced scan

曹锋 1王传彬 1董江宁 1高飞 1李乃玉1

作者信息

  • 1. 安徽省肿瘤医院影像科 安徽 合肥 230031
  • 折叠

摘要

Abstract

Objective: To investigate the difference of image quality and radiation dose between Neusoft NeuViz 128 CT and GE HD750 CT for neck enhanced scan with the similar scan parameters by using comparative study. Methods: The patients whose the range of Body Mass Index (BMI) was 18.5~24.9 Kg/m2were divided into observation group (60 cases who received scan by using Neusoft NeuViz 128 CT) and control group (60 cases who received scan by using GE HD750). And the signal noise ratio (SNR), contrast noise ratio (CNR), standard deviation (SD) of CT value of subcutaneous fat, CT dose index volume (CTDIvol), dose length product (DLP), effective dose (ED) and subjective image quality scores of arteria carotis communis and jugular vein between the two groups were analyzed under the same scan conditions. Results: For the arterial phase images, the SNR, CNR and SD of observation group and control group were (42.70±15.75) and (51.34±15.32), (35.60±14.62) and (42.35±13.66), and (7.64±1.94) and (7.13±2.07), respectively, and the differences of these indicators between the two groups were no significant (t=-1.854, t=-1.607, t=0.833, P>0.05). For the venous phase, the SNR, CNR and SD of observation group and control group were (24.25±9.34) and (27.03±8.54), (16.46±7.48) and (16.63±6.33), and (7.71±1.91) and (7.09±1.56), respectively, and the differences of these indicators between the two groups also were no significant (t=-1.038, t=-0.086, t=1.192, P>0.05). And the difference of CTDIvolbetween observation group (12.4±0.00) and control group (13.52±0.25) were significant (Z=-1.725, P<0.05). And the DLP of the two groups were (373.42±14.00) and (372.73±17.55), respectively, and the ED of the two groups were (2.20±0.08) and (2.19±0.11), respectively. And the differences both of DLP and ED between the two groups were no significant (t=0.17, t=0.593, P>0.05). Besides, the subjective image quality scores of the two groups were (4.57±0.50) and (4.63±0.48), respectively, and there was no significant difference between the two groups (Z=-1.523, P>0.05). Conclusion: The same image quality and same radiation dose of neck enhanced scan can be obtained from Neusoft NeuViz 128 CT and GE HD750 CT under the similar condition.

关键词

颈部增强/CT/图像质量/辐射剂量

Key words

Neck enhancement/Computed tomography/Image quality/Radiation dose

分类

医药卫生

引用本文复制引用

曹锋,王传彬,董江宁,高飞,李乃玉..NeuViz 128型CT颈部增强图像质量与辐射剂量的比较评估[J].中国医学装备,2018,15(5):10-13,4.

基金项目

2016年度国家工业和信息化部高端医疗设备应用示范项目(2016MHD11002002103)"东软NeuViz 128 CT与GE Discovery CT 750HD在肿瘤影像学中的应用对比研究" (2016MHD11002002103)