射波刀放疗中相关螺旋CT定位条件的探讨

陈国梁 赵瑞 李俊杰 李莎

中国医疗设备2018,Vol.33Issue(5):90-92,3.
中国医疗设备2018,Vol.33Issue(5):90-92,3.DOI:10.3969/j.issn.1674-1633.2018.05.024

射波刀放疗中相关螺旋CT定位条件的探讨

Discussion of the Location Conditions of Spiral CT in Cyberknife Radiotherapy

陈国梁 1赵瑞 1李俊杰 1李莎1

作者信息

  • 1. 解放军兰州总医院 放疗科,甘肃 兰州 730050
  • 折叠

摘要

Abstract

Objective To investigate the location conditions of spiral CT in Cyberknife radiotherapy. Methods Siemens big bore 64 slice spiral CT was used to scan the chest tumor phantom with different pitch and reconstruction slice thickness. The tumor phantom volume, length and CT value with the software of Siemens CT workstation was measured and the changes of the tumor volume, length and CT value was observed. Results Different scanning pitch had no effect on tumor volume, tumor length and CT value. The tumor volume and CT value decreased with the increase of the thickness of the reconstructed layer. The tumor size increased with the thickness of the reconstructed layer. When the slice thickness was more than 1.5 mm, the tumor volume was significantly decreased with the thickness of the reconstructed layer. When the reconstruction slice thickness was 5 mm, the tumor volume changeed up to 23%. Conclusion In order to ensure the quality of cyberknife radiotherapy, the CT location condition should be executed strictly in cyberknife locating requirements.

关键词

射波刀/放射治疗/重建层厚/螺距/肿瘤

Key words

cyberknife/radiotherapy/reconstruction slice thickness/pitch/tumor

分类

医药卫生

引用本文复制引用

陈国梁,赵瑞,李俊杰,李莎..射波刀放疗中相关螺旋CT定位条件的探讨[J].中国医疗设备,2018,33(5):90-92,3.