| 注册
首页|期刊导航|中国医疗设备|某地区医疗机构CT性能状态检测结果分析

某地区医疗机构CT性能状态检测结果分析

姚国庆 丁琴 米永巍 郑小溪 李涛 谭先健 丁朝飞

中国医疗设备2018,Vol.33Issue(5):97-99,3.
中国医疗设备2018,Vol.33Issue(5):97-99,3.DOI:10.3969/j.issn.1674-1633.2018.05.026

某地区医疗机构CT性能状态检测结果分析

Analysis of CT Performance Test Results of Medical Institutions in a Region

姚国庆 1丁琴 2米永巍 1郑小溪 1李涛 1谭先健 1丁朝飞3

作者信息

  • 1. 中国人民解放军武汉总医院医学工程科,湖北 武汉 430070
  • 2. 中国人民解放军武汉总医院医学眼科,湖北 武汉 430070
  • 3. 南方医科大学 研究生院,广东 广州 510515
  • 折叠

摘要

Abstract

Objective To understand the quality of clinical application of CT through the analysis of CT performance test results of medical institutions. Methods The main performance indicators of 39 CT selected in 9 medical institutions in a region were tested for quality control according to national and industry standards. Results Most CT performance status were excellent. The noise pass rate was 97.4%, the measurement index pass rate was 92.3%, and the remaining test indicators pass rate reached 100%. Conclusion CT performance testing is the premise and technical guarantee for its quality of clinical application. In order to better ensure the level of performance of medical units at all levels of CT, we need to strengthen the equipment management consciousness, improve the equipment quality control standard specification and share of quality control information, etc.

关键词

CT性能检测/医疗设备质量控制分析/质控标准规范/设备维护方案

Key words

CT performance detection/quality control analysis of medical equipment/quality control standard specification/equipment maintenance program

分类

医药卫生

引用本文复制引用

姚国庆,丁琴,米永巍,郑小溪,李涛,谭先健,丁朝飞..某地区医疗机构CT性能状态检测结果分析[J].中国医疗设备,2018,33(5):97-99,3.

基金项目

湖北省卫生和计划生育委员会科研项目(WJ2017H0032). (WJ2017H0032)