医疗卫生装备2018,Vol.39Issue(5):68-70,74,4.DOI:10.7687/j.issn1003-8868.2018.05.068
2种X线曝光模式下双下肢全长摄影中图像质量和辐射剂量的比较
Comparison of image quality and radiation doses of double lower extremities full-length photography by two X-ray exposure modes
刘新丽 1李天然 1雷振兴 1刘刚 1宾盼 1徐启明1
作者信息
- 1. 解放军总医院第一附属医院放射科,北京100048
- 折叠
摘要
Abstract
Objective To compare two exposure modes on image quality and radiation dose when used for double lower extremities full-length photography so as to provide guidance for clinical application. Methods Totally 60 patients undergoing double lower extremities full-length photography were selected with sequential experiment and then divided into two groups randomly and equally,of whom,there were four ones with genu varum.In the two groups,one group applied the exposure mode of AEC,and the other involved in FIXED mode.Then the two groups were compared on image quality and radiation dose. Results The exposure radiation to the patient under FIXED mode was higher than that under AEC mode although the common patients had the image quality with no significant difference in the two groups, and FIXED mode resulted in higher image quality for the patients with genu varum than AEC mode.Conclusion Exposure modes have effects on image quality and exposure dose,and it's suggested that AEC mode be applied for common patients from the aspect of radiation protection while FIXED mode be used for the patients with genu varum. [Chinese Medical Equipment Journal, 2018,39(5):68-70,74]关键词
X线/曝光模式/双下肢全长摄影/图像质量/辐射剂量Key words
X ray/exposure mode/double lower extremities full-length photography/image quality/radiation dose分类
医药卫生引用本文复制引用
刘新丽,李天然,雷振兴,刘刚,宾盼,徐启明..2种X线曝光模式下双下肢全长摄影中图像质量和辐射剂量的比较[J].医疗卫生装备,2018,39(5):68-70,74,4.