医疗卫生装备2017,Vol.38Issue(10):126-128,3.DOI:10.7687/j.issn1003-8868.2017.10.126
XCY2002-1/200野战X线诊断车机架运动升级改造研究
Upgrades of XCY2002-1/200 field X-ray diagnosis vehicle rack movement
周峰 1张烨 1张昌戎1
作者信息
- 1. 730050兰州,原兰州军区联勤部药品仪器检验所
- 折叠
摘要
Abstract
Objective To upgrade the components with defects to decrease the failure rate of XCY2002-1/200 field X-ray diagnosis vehicle rack movement.Methods The controlling part was replaced without changing rack structure,limit switch,motor,electric wire as well as the functions and operation method.Results Eight-month trial of three vehicles proved there were no failures in rack movement.Conclusion The upgraded rack gains high stability and eliminates most of hidden risks.关键词
X线诊断车/野战/机架运动/改造/故障Key words
X-ray diagnosis vehicle/field/rack movement/upgrading/failure分类
医药卫生引用本文复制引用