医疗卫生装备2017,Vol.38Issue(12):84-87,4.DOI:10.7687/j.issn1003-8868.2017.12.084
Mobetron2000移动式术中放疗系统的物理性能测试
Physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system
摘要
Abstract
Objective To execute physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system.Methods Mobetron 2000 mobile intraoperative radiotherapy system had its indexes measured according to TG51 protocol such as percent depth dose (PDD),output dose,beam energy,field flatness and symmetry,X-ray pollution,dose linearity and repeatability.Three-dimensional water phantom was used for verification and acceptance when the measurement was executed,while the calimator end was tangent to the water surface and the nominal ource-skin distance was 50 cm.Results All the indexes of the system met the desired requirements except 12 MeV beam energy.Conclusion The system can be promoted clinically for safe,reliable and efficient treatment.关键词
放疗系统/物理性能/放射治疗/加速器/剂量Key words
radiotherapy system/physical performance/radiotherapy/accelerator/dose分类
医药卫生引用本文复制引用
周含,沈泽天,沈君姝,袁喜,朱锡旭,李兵..Mobetron2000移动式术中放疗系统的物理性能测试[J].医疗卫生装备,2017,38(12):84-87,4.基金项目
2017年南京总医院课题(2017051) (2017051)