首页|期刊导航|医疗卫生装备|Mobetron2000移动式术中放疗系统的物理性能测试

Mobetron2000移动式术中放疗系统的物理性能测试

周含 沈泽天 沈君姝 袁喜 朱锡旭 李兵

医疗卫生装备2017,Vol.38Issue(12):84-87,4.
医疗卫生装备2017,Vol.38Issue(12):84-87,4.DOI:10.7687/j.issn1003-8868.2017.12.084

Mobetron2000移动式术中放疗系统的物理性能测试

Physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system

周含 1沈泽天 1沈君姝 1袁喜 1朱锡旭 1李兵1

作者信息

  • 1. 210002 南京,南京总医院放疗科
  • 折叠

摘要

Abstract

Objective To execute physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system.Methods Mobetron 2000 mobile intraoperative radiotherapy system had its indexes measured according to TG51 protocol such as percent depth dose (PDD),output dose,beam energy,field flatness and symmetry,X-ray pollution,dose linearity and repeatability.Three-dimensional water phantom was used for verification and acceptance when the measurement was executed,while the calimator end was tangent to the water surface and the nominal ource-skin distance was 50 cm.Results All the indexes of the system met the desired requirements except 12 MeV beam energy.Conclusion The system can be promoted clinically for safe,reliable and efficient treatment.

关键词

放疗系统/物理性能/放射治疗/加速器/剂量

Key words

radiotherapy system/physical performance/radiotherapy/accelerator/dose

分类

医药卫生

引用本文复制引用

周含,沈泽天,沈君姝,袁喜,朱锡旭,李兵..Mobetron2000移动式术中放疗系统的物理性能测试[J].医疗卫生装备,2017,38(12):84-87,4.