中国医疗设备2018,Vol.33Issue(7):22-26,5.DOI:10.3969/j.issn.1674-1633.2018.07.005
国产医用加速器XHA1400的X射线剂量和机械性能测试
X-Ray Dose and Mechanical Properties Test of Domestic Medical Accelerator XHA1400
摘要
Abstract
Objective In order to evaluate the clinical feasibility and reliability of three-dimensional conformal radiotherapy and intensity modulated radiation therapy of domestic middle-energy medical accelerator XHA1400, we examined the X-ray dose output, mechanical property and safety of it. Methods Based on GB15213-2016 Medical Accelerator Performance and Test Method and AAPM TG142, and referring to the related accelerator quality control standards of imported equipment manufacturers such as Elekta, we conducted a comprehensive verification of the accelerator XHA1400. Results In addition to the results of the linear requirements, XHA1400 indicators had reached quality control standards. Conclusion XHA1400 has good dose output accuracy and mechanical performance accuracy, and the main indicators are not inferior to the imported accelerator. It meets the most of clinic requirement of the three-dimensional conformal radiotherapy and intensity modulated radiotherapy. However, there is still some shortages. Its dose output linearity should be further calibrated in order to better meet the requirement of small field dose output in clinical intensive radiation therapy. What's more, we should equip the accelerator with cone-beam CT for higher imaging quality.关键词
国产放疗设备/验收测试/剂量特性/机械性能/质量控制Key words
domestic radiotherapy equipment/acceptance test/dose characteristics/mechanical properties/quality control分类
医药卫生引用本文复制引用
吴青南,高源,王运来,鞠忠建,袁树海,戴相昆,宋孟孟,宋迎新,王云,陈富宝..国产医用加速器XHA1400的X射线剂量和机械性能测试[J].中国医疗设备,2018,33(7):22-26,5.基金项目
国家重点研发计划(2016YFC0105715). (2016YFC0105715)