| 注册
首页|期刊导航|中国医疗设备|飞利浦Brilliance 16排CT故障分析与维修

飞利浦Brilliance 16排CT故障分析与维修

徐旭 陈曦

中国医疗设备2018,Vol.33Issue(7):96-97,115,3.
中国医疗设备2018,Vol.33Issue(7):96-97,115,3.DOI:10.3969/j.issn.1674-1633.2018.07.027

飞利浦Brilliance 16排CT故障分析与维修

Philips Brilliance 16 Row CT Failure Analysis and Maintenance

徐旭 1陈曦2

作者信息

  • 1. 徐州市第一人民医院 设备科,江苏 徐州 221002
  • 2. 徐州市计量检定测试中心,江苏 徐州 221009
  • 折叠

摘要

Abstract

First of all, brief introduction and working principle of philips Brilliance 16 row CT were introduced in the present paper. Then, three consecutive of the philips Brilliance 16 row CT computed tomography that showed on the machine operation interface after more than 10 patients boot scanning were analyzed. The failure codes which including "Gantry can't comply please try again""Patient table cannot comply, please retry" and "Tilt system cannot comply, please retry". These faults still occured after the machine was restarted. Finally, the failures were solved by replacing the R2D board. Because the faults belonged to the ring type fault, it was necessary for the engineer to judge according to the actual maintenance experience and error code, and solve the problem. Through the process of this repair, we hope to provide some assistance to the clinical maintenance engineers during the maintenance.

关键词

Brilliance 16排CT/报错代码/旋转机架/R2D控制板/控制电路/滑动丝杆/设备维修

Key words

brilliance 16 row CT/error code/rotating frame/R2D control board/control circuit/slide rod/equipment maintenance