Objective To ensure the safety of patients with hemodialysis, the accuracy of the parameters between these two types of machines was analyzed through the quality control test of two different types of hemodialysis machines in our hospital. Methods The conductivity, temperature and flow parameters of three 4008S hemodialysis machines and three 5008S hemodialysis machines procurement with the same year were tested respectively using HDM99XP. It was tested once every three months, four times in total, and the measured data were analyzed. Results The parameters of 4008S and 5008S were within the error range. But the three core parameters’s deviation of temperature, conductivity and flow rate of the 4008S were all greater than 5008S, and the parameters with large deviation were corrected in time. Conclusion Stability of dialysis parameters in hemodialysis of 5008S is superior to 4008S. The significant differences in the clinical effect of dialysis on patients of these two types machines may continue to be studied.

关键词

血液透析机/血液透析机质量控制/电导率/透析液温度/透析液流量

Key words

hemodialysis machine/quality control/conductivity/dialysate temperature/dialysate flow

分类

机械制造

引用本文复制引用

刘广全,梁静,常玉锋,吴晓亮..我院两种不同类型血液透析机的质量控制对比及分析[J].中国医疗设备,2018,33(9):56-58,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文