飞利浦DR平板探测器故障三例

Three Cases of Troubleshooting of the Flat Panel Detector for Philips DR

任建峰1

作者信息

  • 1. 山西医科大学第二医院 设备维修部,山西 太原 030001
  • 折叠

摘要

Abstract

Objective By describing the three failures and maintenance processes of the Philips digital radiography system that I encountered, the idea and method of solving three faults of the device flat-panel detector was introduced, which could provided references for our colleagues in the hospital to maintain such devices. Methods Firstly, the fault direction was judged based on the equipment failure phenomenon, error messages and various test methods. Then, the long-term maintenance experience coupled with reference hardware circuit diagram was performed. Finally, the fault point was found out and the corresponding parts was replaced, and the problem was solved smoothly. Results Three failures of flat panel detector part were solved smoothly by self-maintenance, and the loss caused by equipment shutdown was successfully avoided. Conclusion As long as the correct maintenance ideas are mastered and the working principle of the equipment hardware circuit is analyzed, the self-maintenance of this kind of equipment can be achieved, which can achieve the purpose of improving the personal maintenance technology, ensuring the opening rate of the equipment and saving expenses for the hospital.

关键词

数字化拍片系统/平板探测器/设备故障/医疗设备维修

Key words

digital radiography/flat panel detector/equipment failure/medical equipment maintenance

分类

信息技术与安全科学

引用本文复制引用

任建峰..飞利浦DR平板探测器故障三例[J].中国医疗设备,2018,33(9):97-99,3.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文