南通大学学报(自然科学版)2018,Vol.17Issue(4):61-65,5.
离子色谱-安培检测法检测饮用水中氰化物的方法分析
Method Analysis for Determination of Cyanide in Drinking Water by IC-PAD
傅珠婷 1石健 1蔡萌 1李春建1
作者信息
- 1. 南通大学 分析测试中心, 江苏 南通 226019
- 折叠
摘要
Abstract
A method for rapid determination of trace cyanide in drinking water is established using an ion chromatography-amperometric detector. The minimum detection limit of this method is 0.17 μg/L (signal-to-noise ratio S/N =3) , the minimum limit of quantification is 0.56 μg/L (signal-to-noise ratio S/N = 10) , and has a good linearity over a concentration range of 0-150 μg/L, and it is not easy to be affected by the conventional anions, sulfides and thiocyanides in drinking water. The test results align with the results of the gas chromatographic GB method when testing the actual samples. Experiments show that the method has good repeatability, low detection limit, high precision and high accuracy. It is not easily affected by other ions during the detection. It is suitable for the determination of cyanide content in drinking water.关键词
离子色谱法/安培检测法/饮用水/氰化物Key words
ion chromatography (IC)/pulsed amperometric detection (PAD)/drinking water/cyanide分类
化学化工引用本文复制引用
傅珠婷,石健,蔡萌,李春建..离子色谱-安培检测法检测饮用水中氰化物的方法分析[J].南通大学学报(自然科学版),2018,17(4):61-65,5.