Mobius 3D软件在调强放射治疗计划剂量验证中的应用研究

于得全 张琰君 常浩 高宏 屈焕敏

中国医学装备2019,Vol.16Issue(1):39-41,3.
中国医学装备2019,Vol.16Issue(1):39-41,3.DOI:10.3969/J.ISSN.1672-8270.2019.01.011

Mobius 3D软件在调强放射治疗计划剂量验证中的应用研究

Research on the application of Mobius 3D software in dosage verification of IMRT plan

于得全 1张琰君 1常浩 1高宏 1屈焕敏2

作者信息

  • 1. 空军军医大学唐都医院放射治疗科,陕西 西安 710038
  • 2. 空军军医大学唐都医院政治部,陕西 西安 710038
  • 折叠

摘要

Abstract

Objective:To study the application of Mobius 3D intensity-modulated verification software in dosage verification of intensity-modulated radiation therapy (IMRT) plan.Methods:80 patients with tumor (34 cases with head and neck tumor, 33 cases with chest tumor, 13 cases with abdominal tumor) were selected.The corresponding IMRT plan was designed on the Varian Eclipse 11 plan system and PTW1500 Two-Dimensional Ionization Chamber Matrix was used to verify IMRT Plan before patients underwent treatment.And according to the requirements of intensity-modulated verification, the IMRT plan was send to Mobius 3D software workstation, and patients were treated at the accelerator.The Mobius 3D verification system was used to obtain the Dynalog file of therapy in accelerator, and to implement automatic calculation with convolution and (or) superposition so as to generate 3D dose of treatment for target region.And then it was compared with the 3D dose of therapy plan system (TPS) .Besides, Gamma pass rates which limiters were the dosage error (3%) and position error (3mm) (3%/3mm) was adopted to implement analysis.Results:Gamma pass rate (3%/3mm) of IMRT plan of 80 patients with head and neck, with thoracic cancer, abdominal tumor were 98.33±1.67%, 99.02±0.98%and 98.73±1.07%, respectively.And the IMRT plan of all patients could meet the clinical needs.Conclusion:Mobius 3D software meets the need of clinical dose verification of IMRT plan.

关键词

Mobius 3D软件/调强放射治疗计划/剂量验证/Gamma通过率

Key words

Mobius 3D software/Intensity-modulated radiation therapy (IMRT) plan/Dosimetric verification/Gamma pass rate

分类

医药卫生

引用本文复制引用

于得全,张琰君,常浩,高宏,屈焕敏..Mobius 3D软件在调强放射治疗计划剂量验证中的应用研究[J].中国医学装备,2019,16(1):39-41,3.

中国医学装备

OACSTPCD

1672-8270

访问量0
|
下载量0
段落导航相关论文