临床神经外科杂志2019,Vol.16Issue(3):190-193,4.DOI:10.3969/j.issn.1672-7770.2019.03.002
脑深部电刺激术同期植入两侧电极精度的对比研究
Comparative study on accuracy of simultaneous bilateral electrodes implantation in DBS
摘要
Abstract
Objective To compare the accuracy of simultaneous implantated bilateral electrodes in deep brain stimulation ( DBS ).Methods 133 patients with bilateral DBS from February 2017 to February 2019 in the department of neurosurgery , the General Hospital of Northen Theater Command were included.A total of 266 electrodes were implanted , including 160 sides of subthalamic nucleus( STN), 2 sides of ventromedial nucleus of thalamus and 104 sides of globus pallidus interior( Gpi).All patients underwent three-dimensional reconstruction of the head 2 hours after operation and 1 week after operation , which was fused with the preoperative surgical plan.The distance between the implanted electrode and the preoperative target on the X and Y axes in the target plane was measured.Results All 266 electrodes were successfully implanted , and no serious complications such as intracranial hemorrhage , intracranial infection and hemiplegia occurred , and no electrode fracture occurred.There was no statistically significant difference in the error (all P>0.05), between the first implanted electrode and the second implanted electrode on the X -axis and the Y-axis, neither at 2 hours postoperatively nor 1 week after surgery.Conclusion DBS can accurately implant both electrodes at the same time without delaying the implantation of the second electrode .关键词
脑深部电刺激/机器人/电极精度/延期植入Key words
deep brain stimulation/robot/electrode accuracy/delayed implantation分类
医药卫生引用本文复制引用
SUN Xiao,TAO Ying-qun,YANG Xing-wang..脑深部电刺激术同期植入两侧电极精度的对比研究[J].临床神经外科杂志,2019,16(3):190-193,4.基金项目
国家自然科学基金面上项目(81870890 ) (81870890 )
辽宁省自然科学基金计划重点项目(20170540935) (20170540935)