中国医疗设备2024,Vol.39Issue(1):24-28,5.DOI:10.3969/j.issn.1674-1633.2024.01.005
一种7.5F自研心腔内超声动物实验
Animal Experiment of a 7.5F Self-Developed Intracardiac Echocardiography
冯文慧 1何精才 1曹琴 1周欣欢1
作者信息
- 1. 深圳心寰科技有限公司,广东 深圳 518101
- 折叠
摘要
Abstract
Objective To verify the performance and safety of intracardiac echocardiography(ICE)imaging system in animal experiments.Methods The imaging depth and resolution of ICE system were tested by standard ultrasonic phantom,then Bama pig was imaged by ICE,and the imaging field and resolution were evaluated.At the same time,the catheter passability,bending maneuverability and safety were evaluated,and compared with the commercial ICE system.Results The self-developed ICE imaging system had a maximum detection depth of≥60 mm and axial and lateral resolution of≤1.0 mm.In vivo ICE imaging,two kinds of catheters successfully entered the right atrium through femoral vein under the guidance of digital subtraction angiography,and the target section could be displayed by adjusting the angle of the catheter tip with the handle.The structure of the cardiac chamber was clear and recognizable,with less noise,and the imaging quality of the self-developed ICE system was close to that of the commercial ICE system.After the operation,all the experimental pigs survived healthily and no intervention-related injuries were found.Conclusion The self-developed 7.5F ICE catheter system is stable,safe and easy to operate.关键词
心腔内超声/心脏介入手术/相控阵换能器/介入导管Key words
intracardiac echocardiography/cardiac intervention/phase array transducer/intervention catheter分类