| 注册
首页|期刊导航|中国医学装备|不同测试方法在精密手术器械功能质量检查中的比较效果研究

不同测试方法在精密手术器械功能质量检查中的比较效果研究

车凤莲 钱黎明 陈维维 郑乐乐 陈玉 潘慧园

中国医学装备2024,Vol.21Issue(3):129-132,4.
中国医学装备2024,Vol.21Issue(3):129-132,4.DOI:10.3969/j.issn.1672-8270.2024.03.026

不同测试方法在精密手术器械功能质量检查中的比较效果研究

Comparative effectiveness research of different testing methods in functional quality inspection of precision surgical instruments

车凤莲 1钱黎明 2陈维维 3郑乐乐 3陈玉 3潘慧园3

作者信息

  • 1. 上海交通大学医学院附属瑞金医院护理部 上海 200025||上海交通大学医学院附属瑞金医院消毒供应中心 上海 200025
  • 2. 上海交通大学医学院附属瑞金医院消毒供应中心 上海 200025
  • 3. 上海交通大学医学院附属瑞金医院无锡分院消毒供应中心 无锡 214000
  • 折叠

摘要

Abstract

Objective:To compare the testing effects of different testing methods in the functional inspection of precision surgical scissors and needle holders.Methods:By consulting relevant literature and professional books,three testing methods and materials for precision surgical scissors and needle holders were selected.The cutting performance of the precision surgical scissors were tested by gauze cloth,cotton wool,and silicone film method in the Interlock surgical instrument testing set.The clamping performance of precision needle holder was tested by using light,needle clips,and thread clamping method in the Interlock surgical instrument testing set.100 pieces of precision surgical scissors and needle holders which were qualified for cleaning and disinfection in the hospital were selected for effectiveness testing,and a self-made satisfaction questionnaire to investigate the satisfaction of surgeons with the use of instruments.Results:After 100 precision surgical scissors tested using three methods of gauze,cotton wool,and silicone film,the qualified number of cutting performance was 94,76,and 58,respectively,the difference was statistically significant(x2=36.526,P<0.05).The satisfaction rate of surgeons with the use of precision surgical scissors passed the silicone film test was the highest at 100%,followed by cotton wool cloth at 82.0%,and the gauze method at the worst at 67.0%,the difference was statistically significant(x2=48.439,P<0.05).After 100 precision needle holders were tested by using three methods of light irradiation,needle clamping,and wire clamping,the qualified number of clamping performance was 96,78,and 62,respectively,the difference was statistically significant(x2=38.160,P<0.05).The satisfaction rate of surgeons with the precision needle holders passed the clamp line test was the highest at 100.0%,followed by the clamp needle at 79.0%,and the worst under light at 62.0%,the difference was statistically significant(x2=57.705,P<0.05).Conclusion:The function and quality of precision surgical scissors and needle holders tested by the test objects in Interlock Surgical Instrument Test Suite can objectively and accurately reflect the functional status of surgical scissors and needle holders,provide reference for regular maintenance and upkeep of instruments,and ensure safe use of instruments.

关键词

测试方法/精密手术器械/质量控制/器械测试物/切割性能/夹持性能

Key words

Testing methods/Precision surgical instruments/Inspection/Quality control/Instrument test material/Cutting performance/Clamping performance

分类

医药卫生

引用本文复制引用

车凤莲,钱黎明,陈维维,郑乐乐,陈玉,潘慧园..不同测试方法在精密手术器械功能质量检查中的比较效果研究[J].中国医学装备,2024,21(3):129-132,4.

中国医学装备

OACSTPCD

1672-8270

访问量1
|
下载量0
段落导航相关论文