基于半经验法多球中子谱仪校准

Calibration of Bonner Sphere Spectrometer using semi-empirical method

孙博文 1曾志 1张辉 1马豪 1李君利1

作者信息

  • 1. 清华大学工程物理系,北京 100084||清华大学 粒子技术与辐射成像教育部重点实验室,北京 100084
  • 折叠

摘要

Abstract

A new method for calibrating Bonner Sphere Spectrometer by using semi-empirical method to deduct neutron scattering background was proposed.The detection efficiency of the He-3 proportional counters in each neutron sphere of Bonner Sphere Spectrometer was calibrated by using a thermal neutron reference radiation device.The reference 241 Am-Be neutron source was measured separately by each neutron sphere of the Bonner Sphere Spectrometer,and the fluence response of each neutron sphere was given by using the semi-empirical method to deduct the scattering.Combining the results of two experiments,the energy spectrum of 241 Am-Be neutron field was obtained and the fluence calibration factor of the spectrometer was given.The results also verify the ability of calibrated Bonner Sphere Spectrometer to measure fast neutron field.

关键词

多球中子谱仪/校准/中子能谱/半经验法

Key words

Bonner Sphere Spectrometer/calibration/neutron energy spectrum/semi-empirical method