辐射防护2024,Vol.44Issue(2):141-146,6.
测氡仪校准因子的不确定度分析
Uncertainty analysis of calibration factor of radon monitor
摘要
Abstract
The uncertainty of the calibration factor is an important parameter to evaluate the calibration results and is an important basis for the evaluation of the measurement results of radon monitor.In this paper,the uncertainty source of calibration radon monitor is analyzed by using radon chamber.The relative standard uncertainty of the calibration factor was measured.The results show that the main influencing factors of the uncertainty of the calibration factor include:the repeatability of the standard instrument and the measured instrument;standard instrument calibration factor and its annual stability;inhomogeneity of test area of radon chamber device.Finally,the relative standard uncertainties of the calibration factors at each measurement point are 4.1%,3.6%,and 3.3%,respectively.关键词
氡/测氡仪/校准/不确定度分析Key words
radon/radon monitor/calibration/uncertainty analysis分类
能源科技引用本文复制引用
田兴宇,张雄杰,庞柳青,李云洋,苏建文..测氡仪校准因子的不确定度分析[J].辐射防护,2024,44(2):141-146,6.基金项目
中国地震局第一监测中心科技创新主任基金项目(FMC2022004) (FMC2022004)