计量学报2024,Vol.45Issue(4):594-599,6.DOI:10.3969/j.issn.1000-1158.2024.04.20
人体组织液特性对SAR值的影响及修正模型
Impacts of Tissue-equivalent Liquid Properties on SAR and the Correction Model
摘要
Abstract
To determine a correction model that can be used to correct the measured specific absorption rate(SAR)in a homogeneous phantom when its dielectric properties deviates from standardized target values.The correction model is established based on the method of electromagnetic numerical simulation.The method is verified by comparison with the data in the IEC 62209\IEEE 1528 standards at 835 MHz.Based on the method,correction models at 3.5 and 28 GHz,which are the candidate bands for the future communication systems,are proposed.Employing the correction models reduces the SAR measurement uncertainty,thereby improving the reproducibility of SAR compliance assessment between laboratories.Another benefit is that it enables the use of broadband tissue-equivalent liquids,whose dielectric parameters are not currently within the tight tolerances of existing standards.The use of broadband liquids reduces the cost of SAR measurement.关键词
电磁辐射/比吸收率/移动通信/组织液/电磁仿真/SAR值修正模型Key words
electromagnetic radiation/SAR/mobile communication/tissue-equivalent liquid/electromagnetic numerical simulation/revised model of SAR分类