中国医疗设备2024,Vol.39Issue(4):169-173,180,6.DOI:10.3969/j.issn.1674-1633.2024.04.029
光学体表成像设备Catalyst的故障维修案例及日常保养方法
Fault Repair Cases and Daily Maintenance Methods of Optical Surface Imaging Equipment Catalyst
摘要
Abstract
This paper introduced the working principle of optical surface imaging equipment Catalyst HD,as well as methods for dealing with common hardware and software failures,and provided daily maintenance methods.For hardware failures,the MutilZsn software of Catalyst HD system was used to determine the faults of the Catalyst projector and camera.As for software failures,the triggering causes of the common Authorization Pending interlock issues on medical linear accelerators with the Catalyst HD system were discussed.This provided reference suggestions for the department to better implement optical surface-guided radiotherapy techniques and efficiently utilize the equipment.关键词
光学体表引导放射治疗/Catalyst/故障维修/维护保养Key words
optical surface guided radiotherapy/Catalyst/breakdown maintenance/maintenance分类
医药卫生引用本文复制引用
彭旭东,张俞,何垠波,李光俊..光学体表成像设备Catalyst的故障维修案例及日常保养方法[J].中国医疗设备,2024,39(4):169-173,180,6.基金项目
国家自然科学基金项目(81472807) (81472807)