瓦里安Clinac 21EX直线加速器GFIL连锁的故障分析与维修

Analysis and Maintenance of GFIL Interlock Faults in Varian Clinac 21EX Linear Accelerator

袁虎 1杨金泉 2廖廷钊2

作者信息

  • 1. 大理白族自治州人民医院 放疗中心,云南 大理 671000
  • 2. 大理白族自治州人民医院 设备科,云南 大理 671000
  • 折叠

摘要

Abstract

This paper summarized and analyzed the GFIL interlock faults that had occurred in our hospital's Varian Clinac 21EX linear accelerator over the past 9 years.The occurrence of GFIL interlock faults was mainly caused by faults of the power supply board on the cold end,the small power supply board on the hot end,the hot deck filament power supply board on the hot end,the backplane board of the hot end,and patch panel board in the rack.After a fault occurs,the console displayed a red GFIL interlocking alarm,which will continuously flash or remain on.The fault cause and handling method were further analyzed based on the detailed hardware status.The five types of faults described in this paper basically cover most of the hardware related to GFIL interlocking,and provided fast and effective fault solutions based on hardware principles,in order to provide reference for linear accelerator maintenance engineers.

关键词

直线加速器/GFIL连锁/放射治疗/灯丝电源板/电子枪

Key words

linear accelerator/GFIL interlock/radiotherapy/filament power supply/electron gun