| 注册
首页|期刊导航|地震地磁观测与研究|离子色谱仪观测数据影响因素分析

离子色谱仪观测数据影响因素分析

张亚男

地震地磁观测与研究2024,Vol.45Issue(2):130-133,4.
地震地磁观测与研究2024,Vol.45Issue(2):130-133,4.DOI:10.3969/j.issn.1003-3246.2024.02.016

离子色谱仪观测数据影响因素分析

Analysis of influencing factors on observation data of ion chromatography

张亚男1

作者信息

  • 1. 中国北京 100080 北京市地震局
  • 折叠

摘要

Abstract

Based on experiments such as pressure,flow parameters,sample storage time,eluent,and mixed standard solution preparation,this study analyzes the effects of factors such as device system,human interference,site environment,and natural environment on the observation data of ion chromatography.The experimental results show that the flow rate and pressure are inversely proportional to the observed values,and the variation of working current has little effect on the measured values.The purified water brand and template accuracy are important factors in generating errors.Manual operations such as sampling,sample processing,injection,and peak determination in daily observations can cause measurement errors,and it is necessary to ensure the standardization and regulation of observations.

关键词

离子色谱/地震监测/地下流体/标准曲线/淋洗液

Key words

ion chromatography/seismic monitoring/earthquake subsurface fluid/standard curve/eluent

引用本文复制引用

张亚男..离子色谱仪观测数据影响因素分析[J].地震地磁观测与研究,2024,45(2):130-133,4.

基金项目

中国地震局地震监测、预报、科研三结合课题(项目编号:3JH-201901074) (项目编号:3JH-201901074)

中国地震局专项"全国地球物理台网数据跟踪分析与产出"项目 ()

地震地磁观测与研究

1003-3246

访问量0
|