MIL-101(Cr)吸附/热脱附-气相色谱-质谱法检测苯系和氯代VOCs

Determination of Benzene Series and Chlorinated VOCs by MIL-101(Cr)Adsorption/Thermal Desorption-Gas Chromatography and Mass Spectrometry

张玉 1崔进 1梅连瑞 1史芯沂美1

作者信息

  • 1. 天津市产品质量监督检测技术研究院,天津 300384
  • 折叠

摘要

Abstract

A new sampling adsorption tube with MIL-101(Cr)as the filler was constructed,which can be used in thermal desorption-gas chromatography-mass spectrometry for the quantitative detec-tion of 15 typical benzene series and chlorinated volatile organic compounds(VOCs).15 VOCs showed good linear relationship in the ranges of 0-100 ng,with the correlation coefficients(r2)above 0.99.The detection limits of 15 VOCs were 0.7-1.9 ng.The average recoveries of MIL-101(Cr)packing tubes were 86.0%-126%,91.7%-133%and 77.6%-135%at the spiked dosage of 5 ng,30 ng and 70 ng VOC target pollutant,respectively.The relative standard deviations(RSDs)were 6.6%-18%,2.9%-20%and 1.3%-20%,respectively.And the RSDs of the test stability and repeat-ability of MIL-101(Cr)packing tubes were 0.50%-14%and 0.30%-18%,respectively.MIL-101(Cr)packing adsorption tube achieves more efficient adsorption enrichment and thermal desorption of benzene series and chlorinated VOCs based on less mass filler,which helps to reduce the penetration risk of sampled gas.This study expands the practical application scenarios of MOFs materials in the field of environmental monitoring.

关键词

金属有机骨架/MIL-101(Cr)/热脱附-气相色谱-质谱法/吸附管/挥发性有机物

Key words

metal-organic framework/MIL-101(Cr)/thermal desorption-gas chromatography-mass spectrometry/adsorption tube/volatile organic compounds

分类

化学化工

引用本文复制引用

张玉,崔进,梅连瑞,史芯沂美..MIL-101(Cr)吸附/热脱附-气相色谱-质谱法检测苯系和氯代VOCs[J].分析测试学报,2024,43(6):875-882,8.

基金项目

天津市市场监督管理委员会科技计划项目(2021-W02) (2021-W02)

分析测试学报

OA北大核心CSTPCD

1004-4957

访问量4
|
下载量0
段落导航相关论文