不同重建算法对宝石能谱成像定量参数影响的体模研究

贾永军 于楠 张喜荣 于勇 贺太平 段海峰 杨创勃

中国中西医结合影像学杂志2024,Vol.22Issue(4):379-384,389,7.
中国中西医结合影像学杂志2024,Vol.22Issue(4):379-384,389,7.DOI:10.3969/j.issn.1672-0512.2024.04.003

不同重建算法对宝石能谱成像定量参数影响的体模研究

Imaging quality of gemstone spectral imaging with different reconstruction algorithm using a phantom

贾永军 1于楠 1张喜荣 1于勇 1贺太平 1段海峰 1杨创勃1

作者信息

  • 1. 陕西中医药大学附属医院医学影像科,陕西 咸阳 712000
  • 折叠

摘要

Abstract

Objective:To compare the effect of different reconstruction algorithms of gemstone spectral imaging(GSI)parameters,and to explore the feasibility and value of deep learning image reconstruction(DLIR)in quantifying iodine concentration(IC)and effective atomic number(Zeff).Methods:GSI mode was used to scan the QSP phantom which containing 9 test tubes,including 1 tube of pure water and 5 tubes of iodine solution with the iodine concentrations of 3.75,7.5,15,30 and 30 mg/mL.The original scanning data were reconstructed using the standard algorithm ASIR-V(ranged from 0%to 100%with the interval of 10%)and the high/middle/low-strength DLIR(DLIR-L/M/H)with a layer thickness of 0.625 mm for IC and Zeff images.ROIs were placed at the center of the test tubes at 9 fixed layers.The values of IC,Zeff and their average value and standard deviation(SD)of images from different reconstruction algorithms were measured and compared.The values of IC and Zeff under different algorithmic conditions of two tubes of iodine solution with iodine concentration of 30 mg/mL located at the center and edge of the phantom were compared.Results:Different algorithms images were able to be used for the measurements of iodine concentration in pure water and various solutions using IC,and there were no statistically significant differences(all P>0.05).In terms of the SD values of IC,there were no statistically significant differences between DLIR-L and ASIR-V(ranged from 40%to 70%)or between DLIR-M and ASIR-V(ranged from 70%to 100%)of pure water or different iodine solution(all P>0.05),while the remaining SD values showed statistical differences(all P<0.05).The center position IC of the phantom was slightly lower than the edge position,reducing by about 0.63%.Zeff values had no statistically significant differences of different reconstruction algorithms(all P>0.05).In terms of the SD values of Zeff in pure water and iodine solution with iodine concentration of 3.75 mg/mL,there were no statistically significant differences between DLIR-L and ASIR-V(ranged from 40%to 70%)or between DLIR-M and ASIR-V(ranged from 50%to 90%),DLIR-L,DLIR-H(all P>0.05),while the remaining SD values showed statistical differences(all P<0.05).There were no statistically significant differences in the SD values of Zeff of iodine solution with iodine concentration of 7.5,15 or 30 mg/mL in different reconstruction algorithms(all P>0.05).The center position Zeff of the phantom was slightly lower than the edge position,reducing by about 0.13%.Conclusions:Compared to ASIR-V,DLIR algorithm does not change IC and Zeff,but can reduce the noise to some extent,and DLIR-H has the lowest degree of noise.The position change has little effect on IC and Zeff in ASIR-V and DLIR.

关键词

体层摄影术,X线计算机/能谱成像/碘浓度/有效原子序数

Key words

Tomography,X-ray computed/Gemstone spectral imaging/Iodine concentration/Effective atomic number

引用本文复制引用

贾永军,于楠,张喜荣,于勇,贺太平,段海峰,杨创勃..不同重建算法对宝石能谱成像定量参数影响的体模研究[J].中国中西医结合影像学杂志,2024,22(4):379-384,389,7.

基金项目

陕西省教育厅青年创新团队科研计划项目(23JP035) (23JP035)

陕西中医药大学创新团队基金项目(2019-YS04) (2019-YS04)

陕西中医药大学校级科研课题国家基金面上培育项目(2020GP10). (2020GP10)

中国中西医结合影像学杂志

OACSTPCD

1672-0512

访问量0
|
下载量0
段落导航相关论文