现代科学仪器2024,Vol.41Issue(5):166-170,5.
低剂量18F-FDG全身PET/CT肿瘤显像可行性的临床研究
Clinical study on the feasibility of low-dose 18F-FDG whole-body PET/CT tumor imaging
朱丹丹 1何海月 1徐小凯1
作者信息
- 1. 中国科学院合肥肿瘤医院核医学科,合肥 230000
- 折叠
摘要
Abstract
Purpose:To investigate the effect of dose reduction on 18F-FDG total-body PET/CT image quality and lesion detection ability in tumor patients,and to explore the minimum threshold of low dose tracer.Methods:A retrospective analysis was performed on 60 patients with total-body PET/CT examination.Low-dose images were simulated by recodification of list-mode PET data.Image quality,noise and lesion detectability were evaluated using 5-point Likert scales.Measurements of uptake of low-dose images were evaluated using area-of-interest analysis of healthy liver and tumor lesions.Results:My findings showed that the 1/2 and 1/10 groups of the full dose(3.70MBq/kg)had better image quality scores.For small lesions with low SUV uptake,8%and 38%of lesions in the 1/15 dose group and 1/30 dose group were false-negative,respectively.Quantitative analysis of the area of interest showed that SUVmax and SD in liver gradually increased with dose reduction(p<0.05).Conclusion:Total-body PET/CT can significantly reduce the tracer dose to 1/10 of the total dose(0.37MBq/kg).关键词
18-氟脱氧葡糖/全身PET/CT/低剂量/肿瘤Key words
18F-FDG/total-body PET/CT/low-dose/tumor分类
医药卫生引用本文复制引用
朱丹丹,何海月,徐小凯..低剂量18F-FDG全身PET/CT肿瘤显像可行性的临床研究[J].现代科学仪器,2024,41(5):166-170,5.