Simultaneous determination of ethylene oxide,2-chloroethanol and ethylene glycol residues in medical device products by gas chromatography

刘若锦 1刘宝煜 1李挥 1邵文亮 1冯毅1

作者信息

  • 1. 河北省药品医疗器械检验研究院,河北石家庄 050227
  • 折叠

摘要

Abstract

A gas chromatography-based method was developed for the simultaneous and rapid determination of ethylene oxide(EO),2-chloroethanol(ECH),and ethylene glycol(EG)resi-dues in medical devices after EO sterilization.A sample weighing 2.5 g was added with 5 mL of ethanol as the extraction medium,and the residual substances in the sample were extracted at 40 ℃ for 4 h.The samples were separated on a DB-WAX capillary column(30 m×0.53 mm× 1.0μm)and determined using a hydrogen flame ionization detector.The temperature was main-tained at 40 ℃ for 5 min,increased to 120 ℃ at a rate of 40 ℃/min,held for 5 min,and then increased to 200 ℃ at a rate of 6 ℃/min,held for 2 min.The flow rate of the nitrogen gas was 3 mL/min.The split ratio was 5∶1.The inlet and detector temperatures were 200 and 300 ℃,re-spectively.The changes in the chromatographic peak areas over time(0.5-10 h)under differ-ent temperatures(20,30,40,and 50 ℃)were investigated,and the optimal extraction condi-tion was determined to be 40 ℃ for 4 h.In the experiments,quantification was performed using an external standard method.EO,ECH,and EG exhibited good peak shapes and separation effects as well as good linearity within their respective ranges.The linear correlation coeffi-cients for EO,ECH,and EG were greater than 0.99.The limits of detection(LODs)for EO,ECH,and EG were in the range of 0.10-0.40 μg/g,and the limits of quantification(LOQs)were in the range of 0.30-1.20 µg/g.The average recoveries under different spiked levels were in the range of 91.08%-116.08%,and the relative standard deviations(n=6)were in the range of 0.56%-8.45%.EO,ECH,and EG residues were found to exist at different levels in the med-ical devices tested.In particular,disposable infusion sets must be paid careful attention.ECH and EG were not detected in disposable sterile medical devices made of non-polyvinyl chloride materials,which may be due to the fact that the products themselves did not contain chloride ions,they were not exposed to chlorine-containing substances during their production,sterili-zation,storage,transportation,use,etc.This study established a method to detect EO resi-dues in disposable medical devices,and has the advantages of simple operation,excellent spe-cificity,accurate quantification,and good reproducibility.It can simultaneously detect three re-sidual substances in medical devices while meeting the actual detection requirements for EO,ECH,and EG residues.The method can be used to scientifically and effectively evaluate the risk of EO residues in single-use medical devices sterilized with EO,and will be helpful for im-proving the quality of medical devices,ensuring the safety of device use,and providing a refer-ence for regulatory supervision and testing.

关键词

医疗器械/气相色谱法/环氧乙烷/2-氯乙醇/乙二醇

Key words

medical devices/gas chromatography(GC)/ethylene oxide(EO)/2-chloroeth-anol(ECH)/ethylene glycol(EG)

分类

化学化工

引用本文复制引用

刘若锦,刘宝煜,李挥,邵文亮,冯毅..气相色谱法同时测定医疗器械产品中环氧乙烷、2-氯乙醇和乙二醇残留[J].色谱,2024,42(11):1059-1067,9.

基金项目

河北省药品监督管理局科技计划项目(2022ZC1027).Science and Technology Plan Project of the Hebei Drug Administration(No.2022ZC1027). (2022ZC1027)

色谱

OA北大核心CSTPCDMEDLINE

1000-8713

访问量3
|
下载量0
段落导航相关论文