Impact of prosthesis malposition on internal structural strain in unicompartmental knee arthroplasty for osteoporotic patients

依力哈木江·吾斯曼 1刘蒙飞 1巨啸晨 1孙荣鑫1

作者信息

  • 1. 新疆医科大学第六附属医院关节外科,新疆乌鲁木齐,830000
  • 折叠

摘要

Abstract

Objective To investigate the impact of femoral prosthesis malposition on internal structural strain in unicompartmental knee arthroplasty for osteoporotic knees.Methods A normal bone knee joint model was constructed based on imaging studies(CT,magnetic resonance imaging)of healthy volunteers,and an osteoporotic knee joint model was constructed by altering the elastic modulus.Finite element models of femoral prosthesis malposition in unicompartmental knee arthro-plasty were then established for both normal bone and osteoporotic groups.A vertical static load of 1,000 N was applied to both models,and the peak strain values and distributions were observed in the medial compartment structures(polyethylene gasket,cancellous bone under tibial prosthesis)and lateral compartment structures(meniscus,tibial cartilage)at various malposition angles of the femoral prosthesis.Results The peak strain values of the polyethylene gasket,cancellous bone under tibial prosthesis,meniscus,and tibial cartilage in the osteoporotic group were higher than those in the nor-mal bone group at different varus and valgus angles.In the osteoporotic group,the peak strain values of the polyethylene gasket and cancellous bone under tibial prosthesis increased with the increase in fem-oral prosthesis malposition angle,with a greater increase at valgus angles than at varus angles.Similarly,the peak strain values of the meniscus and tibial cartilage increased with the increase in femoral prosthesis malposition angle,but the increase was greater at varus angles than at valgus angles.The strain concentration area of the polyethylene gasket was located in a circular-like area where the fem-oral prosthesis contacted the gasket,the strain concentration area of the cancellous bone was mainly located in the posterolateral region,the strain concentration area of the meniscus was located near the anterior horn,and the strain concentration area of the tibial cartilage was mainly located in the middle near the intercondylar eminence.Conclusion Osteoporosis may adversely affect the strain on internal structures of the knee joint after unicompartmental knee arthroplasty.Varus malposition of the femoral prosthesis may exacerbate the progression of osteoarthritis in the contralateral compart-ment(lateral compartment),while valgus malposition may increase the risk of prosthesis loosening and revision surgery.

关键词

/骨关节炎/骨质疏松/单髁置换术/应变/有限元分析

Key words

knee/osteoarthritis/osteoporosis/unicompartmental knee arthroplasty/strain/finite element analysis

分类

医药卫生

引用本文复制引用

依力哈木江·吾斯曼,刘蒙飞,巨啸晨,孙荣鑫..骨质疏松患者膝关节单髁置换术中假体错位对内部结构应变的影响[J].实用临床医药杂志,2024,28(21):6-12,7.

基金项目

新疆维吾尔自治区自然科学基金资助项目(LFYKYZX2021005) (LFYKYZX2021005)

实用临床医药杂志

OACSTPCD

1672-2353

访问量5
|
下载量0
段落导航相关论文