制冷2024,Vol.43Issue(6):39-45,7.DOI:10.3969/J.ISSN.1005-9180.2024.06.0010
无风管自净型排风柜吸附效率实测分析和研究
Experimental Analysis and Research on Adsorption Efficiency of Ductless Self-filtration Vented Enclosure
摘要
Abstract
The adsorption efficiency of a ductless self-filtration vented enclosure equipped with three sets of parallel adsorption filters was measured,and the conclusion was drawn that the number of heat sources inside the cabinet will greatly affect its filtration efficiency under experimental conditions.At the same evaporation concentration,the adsorption efficiency of three heat sources inside the cabinet is 30%to 50%higher than that of a single heat source when emitting pollutants;When a single heat source emits pollutants from the center of the cabinet,the intermediate filter is the first to be penetrated,resulting in a low effective utilization rate of the entire set of filters and a smaller rated adsorption capacity of the filters.It has been proposed that the current JG/T 385-2012 requires more specific testing methods and evaluation indicators for ductless self-filtration vented enclosures equipped with multiple sets of adsorption filters.关键词
无风管自净型排风柜/吸附过滤器/吸附过滤效率/热源/额定吸附量Key words
Ductless Self-cleaning Exhaust Cabinet/Adsorption Filter/Adsorption Filtration Efficiency/Heat Source/Rated Adsorption Capacity分类
建筑与水利引用本文复制引用