Footscan平台测压系统静态足底压力参数的信度研究

张军 梁颖 苏俊剑

赣南医科大学学报2024,Vol.44Issue(12):1237-1241,5.
赣南医科大学学报2024,Vol.44Issue(12):1237-1241,5.DOI:10.3969/j.issn.1001-5779.2024.12.008

Footscan平台测压系统静态足底压力参数的信度研究

Study on the reliability of static plantar pressure parameters of pressure measurement system on Footscan platform

张军 1梁颖 1苏俊剑1

作者信息

  • 1. 南昌大学附属康复医院,江西 南昌 330006
  • 折叠

摘要

Abstract

Objective:To verify the reliability of the static plantar pressure measurement index in the Footscan platform pressure measurement system,and to provide reference for clinical and research work.Methods:Thirty-eight college students who interned at The Affiliated Rehabilitation Hospital of Nanchang University during November 2021 were selected as the research subjects.Two therapists used the Footscan platform pressure measurement system to conduct three tests on the selected subjects.Therapist A completed the first test X1,and one hour later,therapist B completed the second test Y.After 24 hours,therapist A continued to complete the third test X2.The intraclass correlation coefficient(ICC)was used to analyze the intra-group reliability and inter-group reliability.Results:Among them,the inter-group reliability of static plantar pressure is 0.813(95%CI:1.641-0.903)for the upper left,0.759(95%CI:0.536-0.875)for the lower left,0.746(95%CI:0.512-0.868)for the upper right,and 0.773(95%CI:0.564-0.882)for the lower right.The intra-group reliability of static plantar pressure was as follows:the ICC of the upper left weight in the standing position was 0.722(95%CI:0.468-0.855),the ICC of the lower left weight was 0.664(95%CI:0.357-0.825),the ICC of the upper right weight was 0.665(95%CI:0.358-0.825),and the weight bearing ICC of the lower right side was 0.757(95%CI:0.535-0.873).Conclusion:The static plantar pressure parameters of the Footscan platform pressure measuring system have medium to good reliability.Therefore,it can be used as a reliable basis for clinical research balance and diagnosis of plantar problems.

关键词

Footscan平台测压系统/静态足底压力/组内信度/组间信度

Key words

Footscan platform manometry system/Static plantar pressure/Intra-group reliability/Inter-group reliability

分类

医药卫生

引用本文复制引用

张军,梁颖,苏俊剑..Footscan平台测压系统静态足底压力参数的信度研究[J].赣南医科大学学报,2024,44(12):1237-1241,5.

赣南医科大学学报

1001-5779

访问量0
|
下载量0
段落导航相关论文