ECMO设备血液流量主要影响因素的研究

蔡海青 许朝生 吴少海

中国医疗设备2025,Vol.40Issue(2):29-34,6.
中国医疗设备2025,Vol.40Issue(2):29-34,6.DOI:10.3969/j.issn.1674-1633.20231964

ECMO设备血液流量主要影响因素的研究

Research on the Main Influencing Factors of ECMO Flow Rate

蔡海青 1许朝生 1吴少海1

作者信息

  • 1. 广东省医疗器械质量监督检验所,广东 广州 510663||国家药品监督管理局体外循环器械重点实验室,广东 广州 510663
  • 折叠

摘要

Abstract

Objective To study the main factors affecting the blood flow of ECMO equipment through simulation test.Methods Three experiments were designed.In experiment 1,the blood flow was measured when the pipeline pressure range was 70-130 mmHg at different rotational speeds,and the correlation coefficient between the pipeline pressure and the blood flow was calculated and analyzed.In experiment 2,glycerol/saline mixtures at 1∶3,2∶3 and 3∶3 ratios were used to simulate the viscosity of different tests,and the correlation coefficients between blood flow and viscosity of test media were measured and analyzed.In experiment 3,different test models(patient-side simulation)were designed and compared with experiments 1 and 2 at the same speed.Three experiments were conducted to study the main factors influencing the blood flow of ECMO equipment.Results It was found that the correlation coefficients between the pipeline pressure and the displayed flow value,the measured flow value and the flow error after the static pump were 0.49 to 0.93,0.11 to 0.70 and 0.26 to 0.89,respectively,and the correlation coefficients were strong negative correlation between the glycerol/saline ratio and the flow rate,and the correlation coefficient was-1.000 to-0.961.In addition,different patient-side simulation models also have an impact on blood flow.Conclusion The pressure difference before and after the pump and the viscosity of the test medium are the main factors affecting the blood flow of ECMO equipment,but the pipeline pressure after the static pump has no obvious effect on the blood flow of ECMO equipment.

关键词

ECMO/血液流量/离心泵/心肺转流系统/体外心肺支持辅助设备/管路压力/黏度

Key words

ECMO/blood flow rate/centrifugal pump/cardiopulmonary bypass system/extracorporeal cardiopulmonary support assistive equipment/pipeline pressure/viscosity

分类

医药卫生

引用本文复制引用

蔡海青,许朝生,吴少海..ECMO设备血液流量主要影响因素的研究[J].中国医疗设备,2025,40(2):29-34,6.

基金项目

广东省药品监督管理局科技创新项目(2023ZDZ09). (2023ZDZ09)

中国医疗设备

1674-1633

访问量0
|
下载量0
段落导航相关论文