首页|期刊导航|中国医学装备|SOMATOM Force型双源CT的机械结构分析与常见故障维修

SOMATOM Force型双源CT的机械结构分析与常见故障维修

井赛 汪鹏飞 李洋洋 张洪超 李虎洋

中国医学装备2025,Vol.22Issue(2):175-178,4.
中国医学装备2025,Vol.22Issue(2):175-178,4.DOI:10.3969/j.issn.1672-8270.2025.02.032

SOMATOM Force型双源CT的机械结构分析与常见故障维修

Mechanical structure analysis and common fault repair for SOMATOM Force CT

井赛 1汪鹏飞 1李洋洋 1张洪超 1李虎洋1

作者信息

  • 1. 中国人民解放军联勤保障部队第960医院医学工程科 济南 250031
  • 折叠

摘要

Abstract

Dual-source computed tomography(CT)has a series of functions such as rapid scanning,cardiac imaging,multi-site imaging and spectral imaging,and it has the advantages of high image quality and low radiation dose.The timely maintenance and repair can ensure the stable operation of equipment,and guarantee smooth progress of clinical examinations,and provide reliable basis for the diagnosis and treatment to patients.This paper sorted out and analyzed the mechanical structure about the SOMATOM Force CT equipment from three parts:operation and control,data acquisition,and processing and transmission.It also analyzed the correlation between the mechanical structure and functions around the key components such as X-ray tubes,detectors,collimators,slip rings and high-voltage generators,and clarified the role of each component in CT imaging.In general,this paper summarized the troubleshooting ideas and the key points of maintenance and repair for the mechanical faults of the SOMATOM Force CT scanning bed,cooling device and slip ring,which can accumulate the practical experience in improving the guarantee ability of engineers.

关键词

双源CT/机械故障/冷却装置/维修保养

Key words

Dual-source computed tomography(CT)/Mechanical fault/Cooling device/Maintenance and repair

分类

医药卫生

引用本文复制引用

井赛,汪鹏飞,李洋洋,张洪超,李虎洋..SOMATOM Force型双源CT的机械结构分析与常见故障维修[J].中国医学装备,2025,22(2):175-178,4.

中国医学装备

1672-8270

访问量0
|
下载量0
段落导航相关论文