| 注册
首页|期刊导航|核化学与放射化学|高效液相色谱法同时测定镥[177Lu]氧奥曲肽注射液中两种稳定剂的含量

高效液相色谱法同时测定镥[177Lu]氧奥曲肽注射液中两种稳定剂的含量

王勇 杨柳 李洪玉 梁塑 孔祥瑞 李英 范振亚 罗田伟 王宁 付博 马红利

核化学与放射化学2025,Vol.47Issue(1):86-92,7.
核化学与放射化学2025,Vol.47Issue(1):86-92,7.DOI:10.7538/hhx.2025.47.01.0086

高效液相色谱法同时测定镥[177Lu]氧奥曲肽注射液中两种稳定剂的含量

Simultaneous Determination of Two Stabilizers Content in Lutetium[177Lu]Oxodotreotide Injection by High-Performance Liquid Chromatography

王勇 1杨柳 1李洪玉 2梁塑 1孔祥瑞 1李英 1范振亚 1罗田伟 2王宁 2付博 1马红利1

作者信息

  • 1. 原子高科股份有限公司,北京 102413
  • 2. 原子高科股份有限公司,北京 102413||国家原子能机构核技术(放射性药物工程转化)研发中心,北京 102413||中国核工业集团有限公司放射性药物工程技术研究中心,北京 102413
  • 折叠

摘要

Abstract

A high-performance liquid chromatography(HPLC)was established for the determination of stabilizer(gentisic acid,vitamin C)content in lutetium[177Lu]oxodotreotide injection.The liquid chromatographic parameters were finally confirmed by the optimization experiments of the chromatographic column,mobile phase,elution gradient and UV wavelength.The chromatographic column(4.6 mm×250 mm×5 μm)filled with octadecylsilane bonded organic hybrid silica gel was used,and the mobile phase was 0.82 g/L sodium acetate solution(pH was adjusted to 3.50 with glacial acetic acid)and methanol(gradient elution).The detector system was ultra-violet visible spectroscopic detector(detection wavelength was 295 nm)and radioactive flow detector connnected in series.The column temperature was 25 ℃,the injection volume was 5 μL,the running time was 20.0 min,and the flow rate was 1.00 mL/min with gradient elution.The method can effectively separate gentisic acid and vitamin C,and the resolution meets the requirement of greater than 1.5.The linear ranges of gentisic acid and vitamin C are 0-1.50 g/L and 0-30.0 g/L,respectively.The correlation coefficient r are 1.000 and 0.998,respectively.When the chromatographic conditions of column temperature,flow rate and mobile phase pH are changed,the relative standard deviation(sr)of vitamin C and gentisic acid in the sample solution are 1.62%and 1.16%(n=9)respectively,meet the requirement of less than 2.0%.The sr of the retention and content of gentisic acid in six samples are 0.10%and 0.13%,respectively.The sr of the retention and content of vitamin C in six samples are 0.06%and 0.19%,respectively.The sr of the retention and content of gentisic acid in twelve samples are 0.18%and 0.30%,respectively.The sr of the retention and content of vitamin C in twelve samples are 0.10%and 1.49%,respectively.The above sr results all meet the requirement of less than 2.0%.The rate of recovery of gentisic acid meets the requirement of 90%-1 08%,and the rate of recovery of vitamin C meets the requirement of 92%-105%.This method can accurately determine the contents of gentianic acid and vitamin C without diluting the sample,improving measurement accuracy and reducing radiation dose.The method has good durability,accuracy,and precision.The HPLC method is simple,rapid,sensitive,and accurate,which can be used for the accurate determination of gentisic acid and vitamin C in lutetium[177Lu]oxodotreotide injection,and provide guidance for the optimization of the process of these drugs.

关键词

高效液相色谱法/镥[177Lu]氧奥曲肽注射液/稳定剂/龙胆酸/维生素C/方法验证

Key words

high-performance liquid chromatography/lutetium[177Lu]oxodotreotide injection/stabil-izer/gentisic acid/vitamin C/method validation

分类

化学工程

引用本文复制引用

王勇,杨柳,李洪玉,梁塑,孔祥瑞,李英,范振亚,罗田伟,王宁,付博,马红利..高效液相色谱法同时测定镥[177Lu]氧奥曲肽注射液中两种稳定剂的含量[J].核化学与放射化学,2025,47(1):86-92,7.