新建内镜中心终末漂洗水异常情况分析与对策

赵静 任晓蕊 何子阳 孟祥雨 郭泽 周春莲

中国感染控制杂志2025,Vol.24Issue(3):336-342,7.
中国感染控制杂志2025,Vol.24Issue(3):336-342,7.DOI:10.12138/j.issn.1671-9638.20255462

新建内镜中心终末漂洗水异常情况分析与对策

Analysis and countermeasures for abnormal final rinse water in a newly-built endoscopy center

赵静 1任晓蕊 1何子阳 2孟祥雨 3郭泽 4周春莲1

作者信息

  • 1. 首都医科大学附属北京友谊医院疾病预防控制与感染管理处,北京 101300
  • 2. 首都医科大学附属北京友谊医院消化内镜中心,北京 101300
  • 3. 首都医科大学附属北京友谊医院医学工程处,北京 101300
  • 4. 首都医科大学附属北京友谊医院规划建设处,北京 101300
  • 折叠

摘要

Abstract

Objective To conduct microbiological monitoring before the operation of the purified water treatment system in a newly-built endoscopy center,comprehensively analyze the causes for the standard-exceeding results of microbial detection of final rinse water,propose solutions,and provide reference for handling similar events in the future.Methods Microbial detection data of the final rinse water in the newly-built digestive endoscopy center of a tertiary first-class general hospital in Beijing from April to July 2024 were monitored.The potential causes for standard-exceeding results of microbial detection of final rinse water were analyzed from the perspectives of equip-ment maintenance and management of the purified water treatment system as well as the improvement of cleaning and disinfection methods for the purified water supply pipeline in the endoscopy center,targeted improvement mea-sures were proposed accordingly.Results The microbial monitoring result of final rinse water in the newly-built di-gestive endoscopy center built in April 2024 was 1 400 CFU/100 mL,with the main bacterial type being Cupriavi-duspauculus.After five rounds of improvement measures and rechecks,microbial monitoring result of the final rinse water in the newly-built endoscopy center was 0 CFU/100 mL,with a qualification rate of 100%.Analysis suggested that the main causes for the standard-exceeding results of microbial detection of final rinse water were due to the damage of the reverse osmosis membrane,lack of cleaning for the pure water storage tank before use,and non-standard cleaning and disinfection process for the pure water supply pipeline,after targeted improvement,the problem was solved.Conclusion Medical institutions should continuously conduct periodic monitoring on water used for endoscope,regularly perform cleaning and disinfection of the purified water treatment system,standardize cleaning and disinfection procedures,ensure medical quality and patient safety.

关键词

新建内镜中心/终末漂洗水/微生物/消毒/内镜清洗消毒中心

Key words

newly-built endoscopy center/final rinse water/microorganism/disinfection/endoscope cleaning and disinfection center

分类

预防医学

引用本文复制引用

赵静,任晓蕊,何子阳,孟祥雨,郭泽,周春莲..新建内镜中心终末漂洗水异常情况分析与对策[J].中国感染控制杂志,2025,24(3):336-342,7.

中国感染控制杂志

OA北大核心

1671-9638

访问量2
|
下载量0
段落导航相关论文