贝朗Dialog+血透机故障维修二例

冯美燕 陈健鹏 庞君晖

中国医学工程2025,Vol.33Issue(4):10-12,3.
中国医学工程2025,Vol.33Issue(4):10-12,3.DOI:10.19338/j.issn.1672-2019.2025.04.003

贝朗Dialog+血透机故障维修二例

Two cases of Braun Dialog+hemodialysis machine fault maintenance

冯美燕 1陈健鹏 1庞君晖1

作者信息

  • 1. 广州医科大学附属第一医院,广东 广州 510120
  • 折叠

摘要

Abstract

[Objective]This paper addresses the"Dialysis Access Interference"and"Imminent Repeat DF(Dialysate)Pressure Self-Test"alarms occurring in the B.Braun Dialog+hemodialysis machine,analyzing the faults and proposing solutions to ensure the normal operation of the equipment and the safety of patient treatment.[Methods]By examining the working principles of the hemodialysis machine,a step-by-step troubleshooting approach was adopted to gradually identify the source of the faults.[Results]Both aforementioned faults were resolved,the device passed the self-test,and normal operation was restored.[Conclusion]A single fault may have multiple causes.This study provides a reference method for the rapid diagnosis and repair of similar faults in hemodialysis machines,offering significant practical value for ensuring the smooth implementation of hemodialysis treatment.

关键词

贝朗血透机/故障/维修/透析液通路/压力自检

分类

机械工程

引用本文复制引用

冯美燕,陈健鹏,庞君晖..贝朗Dialog+血透机故障维修二例[J].中国医学工程,2025,33(4):10-12,3.