贝朗Dialog+血透机故障维修二例
Two cases of Braun Dialog+hemodialysis machine fault maintenance
冯美燕 1陈健鹏 1庞君晖1
作者信息
- 1. 广州医科大学附属第一医院,广东 广州 510120
- 折叠
摘要
Abstract
[Objective]This paper addresses the"Dialysis Access Interference"and"Imminent Repeat DF(Dialysate)Pressure Self-Test"alarms occurring in the B.Braun Dialog+hemodialysis machine,analyzing the faults and proposing solutions to ensure the normal operation of the equipment and the safety of patient treatment.[Methods]By examining the working principles of the hemodialysis machine,a step-by-step troubleshooting approach was adopted to gradually identify the source of the faults.[Results]Both aforementioned faults were resolved,the device passed the self-test,and normal operation was restored.[Conclusion]A single fault may have multiple causes.This study provides a reference method for the rapid diagnosis and repair of similar faults in hemodialysis machines,offering significant practical value for ensuring the smooth implementation of hemodialysis treatment.关键词
贝朗血透机/故障/维修/透析液通路/压力自检分类
机械工程引用本文复制引用