Comparison of results obtained from rapid myopia screening device,standard visualacuity chart,and com-puter optometryinstrument in myopia screening

潘红飙 1刘铭 1朱兆锋 1陈珺 1舒娜 1石磊1

作者信息

  • 1. 230041 合肥,安徽省第二人民医院眼科
  • 折叠

摘要

Abstract

Objective By comparing the measurement results of the rapid myopia screening device with the stan-dard visual acuity chart and computer optometry instrument,analyze the differences and correlations between the different methods.Methods Prospective acohort study.Randomly select 4 grades of students from 2 primary schools and 1 junior middle school as research subjects,with 1 class randomly selected for each grade.Use a rapid myopia screening device to measure refractive power and obtain simulated naked eye vision,standard vision table to measure uncorrected vision acu-ity,and computer refractometer to measure refractive power.Compare the differences and correlations between uncorrected vision acuity and equivalent spherical lens power obtained by different methods.Results This study screened a total of 158 students,including 79 males and 79 females,with an average age of(11.15±1.502)years old.There was no statisti-cally significant difference(P>0.05)between the uncorrected vision acuity obtained by the rapid myopia screening instru-ment and the standard vision chart in each grade and overall subgroup.The uncorrected vision acuity results of the two in-struments showed a strong positive correlation(r=0.8717,P<0.05).The equivalent spherical lens power measured by the rapid myopia screening instrument and the computer refractometer showed no statistically significant difference(P>0.05)in each grade and overall subgroup,and the equivalent spherical lens power results of the two instruments showed a strong positive correlation(r=0.8941,P<0.05).Conclusions There was no statistically significant difference in the measure-ment results betweendifferentmethods,and the results were strongly correlated.Both verified the credibility of the measure-ment.Results of this study provide data support for its application in large-scale screening.

关键词

快速近视筛查仪/屈光筛查/电脑验光仪

Key words

Rapid screening device/Refractive screening/computer Optometry instrument

引用本文复制引用

潘红飙,刘铭,朱兆锋,陈珺,舒娜,石磊..快速近视筛查仪与标准视力表、电脑验光仪在近视筛查中的结果比较[J].临床眼科杂志,2025,33(2):164-167,4.

临床眼科杂志

1006-8422

访问量11
|
下载量0
段落导航相关论文