中国标准化Issue(9):252-256,5.DOI:10.3969/j.issn.1002-5944.2025.09.042
胰岛素泵抗扰度试验基本性能验证方法的研究
Research on Validation Method of Basic Performance of Insulin Pumps during Anti-interference Testing
摘要
Abstract
Objective:The aim is to verify the basic performance of insulin pumps under anti-interference environmental conditions,ensuring the safety of users and the effectiveness.Method:By setting a large dose,under interfering environmental conditions,the infusion task was completed within a certain time,and the accuracy of the large dose was verified by weighing the insulin pump before and after infusion.Result:The accuracy of the large dose after interference test is within the range of±5%,and other basic performance meet the requirements.Conclusion:This paper provides a verification scheme and test result analysis method for the basic performance of insulin pump immunity tests,providing guidance and reference for research on electromagnetic compatibility immunity projects.关键词
胰岛素泵/准确性/基本性能/基础率Key words
insulin pump/accuracy/basic performance/basal rate引用本文复制引用
邓振进,谢莉凤,吴碧涛,陈新平,石蓉,周倩,秦婷..胰岛素泵抗扰度试验基本性能验证方法的研究[J].中国标准化,2025,(9):252-256,5.基金项目
本文受湖南省科技厅科药基金项目"胰岛素泵可用性研究"(项目编号:2024JJ8231)资助. (项目编号:2024JJ8231)