| 注册
首页|期刊导航|中国医学装备|德尔格呼吸机故障案例分析与维护保养

德尔格呼吸机故障案例分析与维护保养

郭军 霍英恺 崔宇洋

中国医学装备2025,Vol.22Issue(6):186-188,3.
中国医学装备2025,Vol.22Issue(6):186-188,3.DOI:10.3969/j.issn.1672-8270.2025.06.035

德尔格呼吸机故障案例分析与维护保养

Case analysis and maintenance for malfunction case of Dräger ventilator

郭军 1霍英恺 2崔宇洋1

作者信息

  • 1. 山西医科大学第二医院设备维修科 太原 030001
  • 2. 山西药科职业学院医疗器械维护与管理系 太原 030031
  • 折叠

摘要

Abstract

The cleaning and disinfection,and protection in use of a series of ventilators included Dräger Savina,Evita V and BabylogVN500 were discussed through analyzed the working principle and maintenance case of them.This paper analyzed the cases included the abnormal monitoring of reporting oxygen concentration,the failure of calibration for oxygen concentration,and the abnormal faults of the machine head and communication in using different series of ventilators,and expounded the applications of the one-by-one elimination method and the component substitution method in the maintenance process,so as to effectively locate the fault of equipment,and improve the maintenance efficiency,and provide reference value for the fault handling of the similar equipment.

关键词

工作原理/故障排除/清洁消毒/防护

Key words

Working principle/Troubleshooting/Cleaning and disinfection/Prevention and protection

分类

医药卫生

引用本文复制引用

郭军,霍英恺,崔宇洋..德尔格呼吸机故障案例分析与维护保养[J].中国医学装备,2025,22(6):186-188,3.

基金项目

山西医科大学第二医院基金项目(202203-8) The Second Hospital of Shanxi Medical University Funding Project(202203-8) (202203-8)

中国医学装备

1672-8270

访问量1
|
下载量0
段落导航相关论文