光学体表监测系统结合开放式面罩固定在头部肿瘤放疗摆位中的应用研究

刘鑫业 高丽敏 邓祯祥 许晓云 郭晓钰 樊颐鹏 徐婷婷 谢良喜 张佳佳

临床医学工程2025,Vol.32Issue(7):696-700,5.
临床医学工程2025,Vol.32Issue(7):696-700,5.DOI:10.3969/j.issn.1674-4659.2025.07.0696

光学体表监测系统结合开放式面罩固定在头部肿瘤放疗摆位中的应用研究

Study on the Application of Optical Surface Monitoring System Combined with Open Mask Fixation in the Positioning of Head Tumor Radiotherapy

刘鑫业 1高丽敏 1邓祯祥 1许晓云 1郭晓钰 1樊颐鹏 1徐婷婷 1谢良喜 1张佳佳1

作者信息

  • 1. 厦门大学附属翔安医院肿瘤放疗科,福建厦门 361100||厦门大学医学院,福建厦门 361100||厦门大学,福建厦门 361100
  • 折叠

摘要

Abstract

Objective To study the application effect of optical surface monitoring system(OSMS)combined with open mask fixation in the positioning of head tumor radiotherapy.Methods 34 patients with head tumor who underwent radiotherapy in our hospital from January 2023 to March 2024 were selected as the study subjects.14 patients in group A were positioned with marking line using head thermoplastic closed mask fixation,and 20 patients in group B were positioned without marking line using OSMS combined with head thermoplastic open mask fixation.The patients in group B were further divided into different treatment depth groups for analysis:group B1(SSD=85~90 cm)and group B2(SSD=90 cm and above).In group A,the acquired CBCT images and localized CT images were subjected to bony six-dimensional auto-alignment.In group B,pre-treatment CBCT scans were performed and subjected to bony six-dimensional auto-alignment with localized CT images after adjusting by OSMS positioning.Positioning time consuming and positioning errors in group A and group B,and positioning errors in group B1 and group B2 were compared.Results Positioning time consuming:(95±1.5)s for group A and(95.3±1.6)s for group B,and no significant difference was found between the two groups(P>0.05).Positioning errors:there were statistically significant differences between group A and group B in the Lat,Lng,and Roll directions(P<0.05),but no statistically significant difference was found in the Vrt,Rtn,and Pitch directions between group A and group B(P>0.05);There were statistically significant differences between group B1 and group B2 in the Vrt and Rtn directions(P<0.05),but no statistically significant difference was found in the Lat,Lng,Pitch,and Roll directions between group B1 and group B2(P>0.05).Conclusions OSMS combined with open mask fixation can not only reduce the positioning errors of patients with head tumor radiotherapy,but also improve the comfort of patients;For head tumors of different depths,OSMS monitoring positioning is also reliable.

关键词

头部肿瘤/光学体表监测系统/摆位误差/不同深度

Key words

Head tumor/Optical surface monitoring system/Positioning error/Different depths

分类

医药卫生

引用本文复制引用

刘鑫业,高丽敏,邓祯祥,许晓云,郭晓钰,樊颐鹏,徐婷婷,谢良喜,张佳佳..光学体表监测系统结合开放式面罩固定在头部肿瘤放疗摆位中的应用研究[J].临床医学工程,2025,32(7):696-700,5.

基金项目

2024年厦门市医工结合指导性项目(项目编号:3502Z20244ZD2028) (项目编号:3502Z20244ZD2028)

厦门市卫生健康高质量发展科技计划项目(青年科研课题)(项目编号:2024GZL-QN136) (青年科研课题)

临床医学工程

1674-4659

访问量0
|
下载量0
段落导航相关论文