基于MPPG 5.a的国产TAICHI加速器TPS模型验收与调试

汪之群 雷琪琪 王涛 庞廷田 祝鹤龄 邱杰 杨波

中国医疗设备2025,Vol.40Issue(7):40-48,9.
中国医疗设备2025,Vol.40Issue(7):40-48,9.DOI:10.3969/j.issn.1674-1633.20240507

基于MPPG 5.a的国产TAICHI加速器TPS模型验收与调试

Acceptance and Debugging of the TPS Model of the Domestic TAICHI Accelerator Based on MPPG 5.a

汪之群 1雷琪琪 2王涛 2庞廷田 1祝鹤龄 1邱杰 1杨波1

作者信息

  • 1. 中国医学科学院北京协和医院 放疗科,北京 100032
  • 2. 西安大医集团股份有限公司,陕西 西安 710018
  • 折叠

摘要

Abstract

Objective To verify the accuracy of dose calculation of the beam model established by the domestic TAICHI equipment in the RayStation treatment planning system,so that the equipment can better serve the treatment of clinical patients.Methods Based on the reports of MPPG 5.a and TG119 released by the American Association of Medical Physics,the model dose calculation model of the TAICHI equipment was verified,including PDD curves,Profile curves,TG119,and clinical case verification,etc.Results The gamma pass rates of 2 mm/2%from the basal field of different source skins at shallow depths(0-20 cm)were basically close to 100%.The results at deep depths(25 cm underwater)were worse than those at shallow depths,but the mean value was still above 93%.The dose deviation results at all points were less than 2%(after small field correction).Therefore,the beam current in the model was basically consistent with the actual output of the equipment.The results of gamma pass rates at both point doses and area doses of intensity modulated radiation therapy(IMRT)and volumetric modulated arc therapy(VMAT)for TG119 test cases were much better than the standards required by the TG119 report.The average deviation results of planned point dose verification for clinical patients in IMRT and VMAT were 0.48%and 0.47%respectively,and the gamma pass rates of 2 mm/2%and 3 mm/3%in surface dose tests were 95.58%,99.46%and 96.75%,99.30%respectively.All the results met the clinical needs.The end-to-end lung tests of IMRT and VMAT passed,and the point dose deviations were-0.495%and-2.602%respectively.Conclusion After being verified by the MPPG 5.a and TG119 systems,the results of domestic TAICHI meet and far exceed the reporting requirements and standards,conform to clinical requirements,and can be used to optimize the treatment plans of clinical patients.For cases with poor deep fitting,more attention is needed in the subsequent modeling process.

关键词

TAICHI加速器/TPS模型调试/MPPG 5.a/TG119/国产加速器/调强放疗(IMRT)/容积旋转调强放疗(VMAT)

Key words

TAICHI accelerator/TPS model debugging/MPPG 5.a/TG119/domestic accelerator/intensity modulated radiation therapy/volumetric modulated arc therapy

分类

医药卫生

引用本文复制引用

汪之群,雷琪琪,王涛,庞廷田,祝鹤龄,邱杰,杨波..基于MPPG 5.a的国产TAICHI加速器TPS模型验收与调试[J].中国医疗设备,2025,40(7):40-48,9.

基金项目

中央高水平医院临床科研业务费资助(2022-PUMCH-B-116). (2022-PUMCH-B-116)

中国医疗设备

1674-1633

访问量1
|
下载量0
段落导航相关论文